Safety Instructions

The Tyson Bio HS200-B Blood Glucose Monitoring System is working properly only if the Control Solution test result
is within the specied range printed on the test strip vial. Please repeat the test when the test result is out of range.
Out of range result may be caused by:
˙Incorrect steps while performing the test
˙The Control Solution temperature is lower than 59°F (15°C) or higher than 95°F (35°C)
˙Expired or contaminated Control Solution
˙Expired or contaminated test strips
˙Meter malfunction
Note:
˙The result will not be included in any day average calculation when the Tyson Bio HS200-B Blood Glucose
meter is set in the Control Solution test mode.
˙DO NOT use Tyson Bio HS200-B Blood Glucose meter if the problem persists. Please contact customer service
immediately.
Control Solution Results
25