Safety Instructions

NOTE:
The hypoglycemic (LO) and hyperglycemic (HI) alarm glucose values should be selected in consultation with
your health care provider.
Setting Up HI/LO Alarms
1. Press to enter the HI alarm setting after completing the reminder alarm setup.
2. Press or to set “ON" and press to conrm. If "OFF" is chosen, the meter will skip this setting (Fig.16,17).
3. To change the alarm setting, press or until you reach your desired HI setting and then press to conrm.
(Fig.18) The HI alarm value range is 100 ~ 400mg/dL and the default value is 180mg/dL.
4. Continue to the LO alarm setting (Fig.19), repeat Step 2 and 3 (Fig.20,21) to conrm or to change LO alarm setting.
The LO alarm value range is 40 ~ 90mg/dL and the default value is 70mg/dL.
5. Press to nish setting and the meter will be turned o.
(Fig.16) (Fig.17) (Fig.18)
(Fig.19) (Fig.20) (Fig.21)
12