Product Manual

47
Corrections
31.1
1. a. Turn the Multi-purpose valve to
EXH. & DRY.
b. Turn the handle of the Door Closing
Device ΒΌ of a turn in the clockwise
(closing direction). The pin will be
released, enabling the operator to
open the door.
2. If this does not correct the problem, a
Technician will need to be called
32.1
a. A Tuttnauer Pouch Rack is
recommended to provide proper spacing
of bagged instruments. This will allow
for better exposure to steam for
sterilization and better circulation
of air for drying.
b. If a Pouch Rack is not used, make sure
the instruments on the tray are only one
level deep and that the paper / plastic
bags are plastic side down.
32.2 End cycle according to para. 7.
Possible Cause Checkup and Tests
31.1 Locking pin is trapped.
32.1 Bagged instruments are placed too close to
each other.
32.2 Cycle ending has not been performed
correctly.
Symptom
31. Door handle cannot
be turned counter-clockwise for
the
door to be opened
and there is no pressure in the
chamber.
32. Wrapped or bagged instruments
are not drying properly (see also
sec 7.11).