Product Manual

45
Corrections
16.1 Reset the red lever of the circuit breaker
to the up position. If the problem persists,
have a technician check the unit.
17.1 Open drain valve completely and blow
compressed air into the valve. This should
blow the clog back into the reservoir
where it can be removed. Cover the
reservoir opening to avoid splatter.
18.1 Open drain valve completely and blow
compressed air into the valve. This should
blow the clog back into the reservoir
where it can be removed. Cover the
reservoir opening to avoid splatter.
18.2 Have a technician replace the o-rings.
19.1 Follow cleaning instructions using
Chamber Brite, section 9.
20.1 Use 100% steam distilled water.
20.2 Follow cleaning instructions using
Chamber Brite, section 9.
21.1 Have a technician replace the door bellows.
22.1 Tighten the door more.
22.2 Replace the door gasket (see sec. 8.4).
Possible Cause Checkup and Tests
16.1 A power surge or spike.
17.1 Drain valve is clogged.
18.1 Debris is stopping the valve from seating
properly.
18.2 Drain valve o-rings are worn.
19.1 Unit has not been cleaned recently.
20.1 Not using distilled water.
20.2 Unit has not been cleaned recently.
21.1 The door bellows is leaking.
22.1 Door may not be tightened down enough.
22.2 Door gasket may be worn or cracked.
Symptom
16. Circuit breaker trips.
17. Reservoir does not drain.
18. Drain valve is leaking.
19. Chamber is black or dirty
looking.
20. Chamber has a water mark
along bottom.
21. Steam escaping from the closing
device area.
22. Steam leaking from the door
seal.