Product Manual

41
Corrections
4.1 Have a technician replace the “Dry” light.
5.1 Turn the multi-purpose valve to the DRY
position.
5.2 Turn the timer on.
5.3 Have a technician check out the unit.
6.1 Fill the reservoir with distilled water.
6.2 Follow the instructions in para. 8.6
the valve.
6.3 Follow the instructions in para. 8.6
the piping.
7.1 Clean the air jet, see para. 8.3
7.2 Adjust loading according to "Maximum
Load Sizes" table in sec. 2.
7.3 Items to be sterilized should be separated
and not stacked. See preparations for
sterilization para
is recommended
7.4 Check table para 7.
7.5 See para. 10.9 “unit does not reach the
proper operating pressure”.
Possible Cause Checkup and Tests
4.1 The “Dry” light is burned out.
5.1 The multi-purpose valve is not in the DRY
position.
5.2 The timer is off.
5.3 There is an internal problem.
6.1 There is no water in the reservoir.
6.2 The multi-purpose valve is clogged.
6.3 The fill piping is clogged.
7.1 Air jet is clogged.
7.2 Unit may be overloaded.
7.3 The load may be too densely packed.
7.4 Sterilization time may not be correct.
7.5 Unit not reaching the correct sterilization
pressure reading on the gauge.
Symptom
4. Dry indicator light does not
light up at the beginning of the
dry cycle. The Power light is on
and the unit does heat up.
5. Dry indicator light does not
light up at the beginning of the
dry cycle. The Power light is on
and the unit does NOT heat up.
6. Water does not enter the
Chamber when the multi-
purpose valve is in the FILL
WATER position.
7. Spore tests are failing or
indicator strips are not changing
color according to instructions
of indicator's manufacturer.