Product Manual

40
Corrections
1.1 Make sure the power cord is plugged into
the wall outlet and the back of the unit or
restore power to the wall outlet.
1.2 Reset the circuit breaker.
1.3 Reset the cut out thermostat.
1.4 Have a technician replace the “Power”
light.
1.5 Have a technician checkout the unit.
2.1 Have a technician replace the “Heat” light.
3.1 Turn the multi-purpose valve to the
STERILIZE position.
3.2 Close the door tightly.
3.3 Replace or readjust the door switch
activator. When adjusting – turn the
activator screw, in or out, by ¼ turn until
the door microswitch is pushed in when the
door is closed.
3.4 Turn the timer on.
3.5 Have a technician checkout the unit.
Possible Cause Checkup and Tests
1.1 The unit is not plugged in or the wall outlet
has no power.
1.2 Circuit breaker is tripped.
1.3 Cut out thermostat is tripped.
1.4 Turn
the
multi
-
purpose
valve
to
the
“
Exh
/
Dry
”
position
and
set
the
timer
to
15
minutes.
If the “Dry” light is on, the “Power” light is
burned out. When finished, turn the timer
back to 0 minutes.
1.5 If the “Dry” light does not come on, there is
an internal electrical problem.
2.1 The “Heat” light is burned out.
3.1 The multi-purpose valve is not in the
STERILIZE position.
3.2 The door is not closed tightly.
3.3 The door switch activator is missing or not
adjusted correctly.
3.4 The timer is off.
3.5 There is an internal problem.
10 TROUBLESHOOTING
This troubleshooting section is for use by the Operator of the autoclave. If a problem develops that is not covered in this section, a
qualified technician needs to be called. Please call your dealer or Tuttnauer USA Co.
Symptom
1. Power indicator light does not
light up when the ON / OFF
switch is in the ON position.
2. Heat indicator light does not
light up at the beginning of the
sterile cycle. The Power light is
on and the unit does heat up.
3. Heat indicator light does not
light up at the beginning of the
sterile cycle. The Power light is
on and the unit does NOT heat
up.