10. Remove the tray holder; wipe the interior of the chamber with a damp
cloth.
11. Fill the reservoir with distilled water only.
12. Wipe the tray holder with a damp cloth and return it to the chamber.
13. Turn fill knob to fill position and allow a small amount of water (2-4
ounces) to fill chamber. Remove water from chamber.
14. The autoclave is ready to use.
IMPORTANT!
DO NOT sterilize instruments during the cleaning process!!!
CAUTION!
Keep out of reach of children. Contains mildly acidic ingredients. Avoid
contact with the skin, eyes or clothing. Wash hands well after touching the
powder, in the case of eye contact flush with continuous running water for
at least 15 minutes. If irritation persists get medical attention. If
accidentally swallowed, do not induce vomiting, drink large amounts of
water and obtain medical attention. MSDS available upon request.
For models 1730, 2340, 2540 use one packet of CHAMBER BRITE ™.
For models 3140, 3850, 3870 use two packets of CHAMBER BRITE ™.
Clean every 20 cycles or as needed.
* Total cycle time for cleaning Tuttnauer “M” series is 30 minutes at 273°F
(134°C). Total cycle time for cleaning Tuttnauer “MK” series is 15
minutes at 273°F (134°C). All cycles referenced are from a cold start.