Maintenance/Battery Replacement
Maintenance:
To improve performance and avoid transferring soil borne diseases, clean the sensor tip
by rinsing in alcohol for 5 - 10 minutes.
Replace all batteries if low battery indicator appears, or if readings are faint or unstable.
Store the probe sensor dry.
Important: Do not touch the sensor tip with your fingers. The oils on the skin will
affect to probe’s measurement accuracy.
Changing Batteries:
1. Open battery compartment lid (located
near the LCD screen).
2. Remove and replace the old batteries.
Note polarity shown in battery
compartment. When inserting new
batteries, take care to place them over
the white removal tab. This will make
future battery removal much easier.
3. Recalibrate meter after every battery
change.
Specifications
Readout LCD digital display
EC Range 0.00 - 19.99 mS/cm
Temperature Range -5 - 55 C
Accuracy EC: +/-2% Full Scale
Temperature +/0.5 C
EC Resolution 0.01 mS/cm
Operating Environment 0 to 50C
Calibration 1-point with 2.76 mS/cm standard
Temperature Compensation Automatic form 0 to 50C
Power Four LR44 1.5V alkaline batteries