User's Manual

10/15
Auto power off 3 minute after not being used
The pressure of the running 70KPa~106KPa
Storage Environment. -20~60,10%~95%RH
Operating Environment. 10~40,30%~85%RH
Dimensions 124×95×52mm
Net Weight 236g ( batteries excluded)
Used Life Span of Battery 1000 times
Software version
Use Period 5 year
Application part Out of contact the body,:onnect tubshell
contact the body :cuff
Cuff Dimensions For arm circumference of 22~34cm(Special size to be
ordered separately)
Package Content Cuff ,Instruction Manual,Battery Set ,Warrant Card
Note:Subject to modification without prior notice.
9. EMCDeclaration
Guidance and manufacturer’s declaration – electromagnetic immunity
The XXX” is intended for use in the electromagnetic environment specified below. The customer
or the user of the “XXX” should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.