Blood Pressure Monitor Model:DB68 User Manual Rev.00 2019/03/14 Thank you for purchasing TRULY Blood Pressure Monitor. Please read through this "User’s Manual" before using the product.
Contents 1. Identification For Use ----------------------------------------------------------3 2. Safety Precautions -----------------------------------------------------------3 3. Quick start Guide 3.1 3.2 Monitor -------------------------------------------------------------5 Battery Installation ----------------------------------------------------------6 4. Taking Measurement 4.1 Sitting Correctly ------------------------------------------------------------6 4.
1. Identification For use Truly Automatic Arm Blood Pressure Monitor intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The devices’ features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected.
- When measurement, The pressurization of the cuff can temporarily cause function of simultaneously used monitoring other device on the same limb; The pulse display is not suitable for checking the frequency of heart pacemakers; Do not use a cellular phone near the device, It may result In an operational failure; Do not use the monitor in the driving or flying vehicles; Frequenly measuring will make the blood not flow, cause the arm-numbness and abnormal reaction of body。 When you are using the monitor to
¾ Hand dling batteries properrly lithium-lon batteries are a subjectt to disposa al and Reccycling regulations that vary by c country an nd region.A Always che eck and fo ollow your applicable e regulations before d disposing o any batte of ery contacct recharge eable battery recycling corporattion。 Do D not disa assemble,C Crush,or puncture a battery. Do D not sho ort the exte ernal contacts on a ba attery. Do D not disp pose of a battery b in fiire or wate er.
Package e Contents s 3.2 Batttery Insta allation 1 Pull and a lift on the t lock to remove th he battery cover. c 2 Remo ove the ussed batterie es and inse ert new batteries into the battery b com mpartmentt an shown n,taking carre that the polarities(+ p +) and (-) match m the polarities p of the ba attery comp partment as a indicated d. 3 Repla ace the ba attery coverr. Press it lightly until u the lo ock securrely clicks into place e.
4.2 Applying the e Arm Cufff 1 Place e your elbo ow on a tab ble and palm facing upward,and u d wrap the arm cuff around a the upper lefft arm. 2 Pull on o the end d of the cufff until it wraps w securely around d your upp per arm.Do o not over tighten th he cuff .Allow 1~2 cm m or ½ incch between n the botto om of the ccuff and yo our elbow joint. 3 Make e sure the cuff c is at th he same le evel as you ur heart. Note: --Proper cuff size is critical for accura ate measu urements.
5 . Mem mory Fun nctions After me easuremen nt,the last set of measured values v will be saved d automatic cally after measure ement when the unit is i turned off o by press sing “⊙”bu utton or by automatic power off after 3 minutes. m 5.1 Vie ew Past Re eadings 1 Presss the “MEM M”button to o enter memory functtion, 2 Presss the “MEM M”button to o display th he next setts of value es.Past rea adings are displayed from the most rece ent to the oldest.
polarities. “Er P” displayed Fail to inflate “Er 1” displayed “Er 2” displayed Deflate too rapidly Movement during measurement Signal interference “Er 3” displayed Incorrect results Check if the cuff is properly connected.Replace the air tube if it is broken. Return for servicing. Do not move during measurement Remove interfering source e,g. mobile phones,magnets. Measure again Note: If your problem cannot be solved by the above,consult your store of purchase.Do not disassemble the unit. 7.
Auto power off 3 minute after not being used The pressure of the running 70KPa~106KPa Storage Environment. -20℃~60℃,10%~95%RH Operating 10℃~40℃,30%~85%RH Environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 Power frequency (50/60 Hz) magnetic field <5 % UT (>95 % dip in UT) for 0,5 cycle <5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec <5 % UT (>95 % dip in UT) for 5 sec 3 A/m 3 A/m Mains power quality should be that of a
EMC Declaration (Continued) Guidance and manufacturer’s declaration – electromagnetic immunity The “XXX” is intended for use in the electromagnetic environment specified below. The customer or the user of the “XXX” should ensure that it is used in such an environment.
EMC Declaration Guidance and manufacturer’s declaration – electromagnetic emissions The “XXX” is intended for use in the electromagnetic environment specified below. The customer or the user of the “XXX” should ensure that it is used in such an environment. Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Compliance Group 1 Electromagnetic environment – guidance The “XXX” uses RF energy only for its internal function.
11. Disposal At the end of its use don't dispose the appliance, including removable parts and Accessories, together with the other urban waste, but conformably to 2002/96/EC. Since to be treated apart from home waste, you must take it to a differential collection centre specific for electric and electronic equipment. Otherwise you can give it back to the retailer as you buy an equivalent apparatus.