Full Product Manual

Why?
Microprocessor is fail
Solution
Contact distributor for after service
E33
Why?
In humidity display of model 3&4 only,
means humidity measuring is fail
Solution
Contact distributor for after service
Why red LED is flashing but recorded data is OK
Why?
Alarm detection works every 10 seconds but
data logging are possibly programmed as
every 2 hours. In this condition, if alarm
happens between two logging points, it
might lead the recorded data is OK but red
LED is flashing, especially if the alarm mode
is set as Single.
Solution
To program the alarm function as
cumulative can reduce this situation