Weight ~90g
Battery 2PC 3.0V CR2032
Sampling Interval 30 seconds, 5, 10, 30, 60, 90, 120 minutes
Start delay 0, 5, 30, 45, 60, 90,120 minutes, 24 hours
Alarm range Temp: -30.0~70.0
o
C Humidity:1~99%rH
Baro:300~1100hpa
Alarm delay 0, 5, 30, 45, 60, 90, 120 minutes
Alarm type Single, Cumulative, Disable
Operation keys 3 Keys, Start/Stop & Mark & MX/Mn
LED indicator REC, High /Low alarm
Protection class
Model 2: IP65
Directives EN12830
Operating System Windows only
Error Codes
- - - - . -
Why?
In model 2 only.
External probe is unplugged
Solution
Check the contact between probe and logger
E02
Why?
Measured value is below specified range
Solution
Put the logger in specified range
E03
Why?
Measured value is above specified range
Solution
Put the logger in specified range
E04
Why?
In humidity display of model 3 & 4 only,
means temperature is in error mode
Solution
Solve the error issue of temp. sensor
E11
Why?
In humidity display of model 3 &4 only,
means humidity calibration is fail
Solution
Re-calibrate the humidity
E31