Full Product Manual

- Unit of Measurement (UoM)
- Select the unit that will be displayed on the PDF report and LCD. The
selections are Metric or Imperial. In Metric, it is Celsius for temperature and
hpa for pressure. In Imperial, it is Fahrenheit for temperature and inHg for
pressure.
- Password
- The Password function is default OFF. The user may enable it to prevent
unauthorized reprogramming prior to pressing the START button. A
Password may have up to 16 alpha or numeric characters.
- Company Name
- A user defined name, or descriptor, can be input under Company Name. It
will be displayed on the PDF report as the Title, with a maximum of 20
characters.
- Alarm types
- Regardless of the Alarm Type, if the red LED is triggered, it won’t stop even
the reading return to normal range or logging is stopped. The way to stop
red alarm LED is to plug logger into computer to generate report or power
off the logger.
- Single: an alarm is triggered immediately when the measured value exceeds
the alarm threshold.
- Cumulative: an alarm is not triggered when the measured value exceeds the
alarm threshold, but only once the overall average value during alarm delay
duration exceeds the alarm threshold.
- Disable: No alarm function during the logging process
- Alarm delay
- The preset alarm delay interval for a single alarm type is always ZERO.
- The adjustable alarm delay interval for a cumulative alarm type can be 5
minutes to 2 hours.
- Alarm Limits
- Select the alarm threshold values. For example: if 2 - 8 deg C is selected, it
means that an alarm condition will occur below 2 deg C, or above 8 deg C.
The programmable alarm limit of each parameter is limited to one decimal.
To summarize, to activate a alarm when 30 min average value of hazardous
alarm delay period is higher than 8, please program the alarm delay as 30
mins, alarm type as cumulative and range as 2~8
- Time zone