Full Product Manual

Operation
NOTES:
- Use a coin to twist battery compartment to the UNLOCK position and Install
(2) CR2032 Batteries observing polarity.
- Adobe Reader Software is required
- Program and generate the reports at Ambient Room Temperature. For
example, if removed from a freezer allow stabilization time to room temp
before generating the reports.
Step1 Configuring the data logger
- It is possible to make configuration changes at any time before the logging
function is started. Once the logger is started, configuration changes cannot
be made unless stopping the logging first. If the logger is locked with a
Password, the Password is required to make configuration changes.
- - Press ON/OFF to power on the logger
- - Connect the data logger to a PC via the USB port. The green LED is ON
while building the connection with computer.
- - Windows file display window will appear.
- - If the correct file folder is not open, click on Folder to view files.
- - Open the file " PDF Logger Configuration Tool. exe".
- - The default language is English. The user may change it to one of several
alternate languages. There are English, German, French, Italian, Spanish and
Portuguese.
- - To review the instruction manual, you may click "Manual" to open the
manual PDF file.
- - The user can check the calibration report for the logger by selecting the
PDF file named "Calibration Certificate".
- - User programmable parameters are as follows:
- Sampling rate
- Select the sampling interval you need from 30 seconds to 2 hours.
- Start delay
- Select the start delay from 0 min to 24 hours. For example: If the delay is 5
minutes, and the sampling rate is 10 minutes, the real time to log the first
temperature measurement is 5 minutes after the START button is pressed.
All measurements after the first measurement will be at a 10 minute (or
selected) interval.