Storage
Temperature
-10 to 60
o
C (14 to
140
o
F)
Operating Humidity
<80% RH
Storage Humidity
<80% RH
Operating Altitude
Battery
Battery life
80 hours approx. (if Backlight and Laser are used
continuously the
Battery current
battery life is reduced to 2 to 3 hours approx.)
8.3 mA DC approx.
Weight
Dimensions
Main instrument: 178 x 74 x 33mm (7.0 x 2.9 x1.2")
Infrared Measurement Considerations
When taking IR measurements the meter automatically
compensates for ambient temperature changes. Note that it may
take up to 30minutes to adjust to extremely wide ambient
changes.
Low temperature measurements quickly followed by high
temperature measurements may require several minutes to
stabilize as a result of the IR sensor cooling process.
If the surface of the object under test is covered with frost, oil,
grime, etc., clean before taking measurements.
If an object's surface is highly reflective apply masking tape or flat
black paint before measuring. Steam, dust, smoke, etc. can obstruct
measurements.
To find a hot spot, aim the meter outside the area of interest then
scan across (in an up and down motion) until the hot spot is
located.