A-6
Appendix For Scale Technicians
PARAMETER
FUNCTION
OPTIONS
C05
Zero Calibration
0 = skip zero calibration
1 = proceed with zero
calibration
Description:
With C05 flashing, press the
key, then use the up
arrow key to change the 0
parameter setting to 1.
Press the key to
proceed. The zero cal will
count down from 10.
Press the key to
proceed to the next
parameter.