Operating Guide TS-700 Series Digital Indicators Includes Appendix for Scale Technicians Set Up • Calibration • Connectivity TS 700-MS Series 2 TS 700-SS Series 2 TS 700-WB Compatible with DataLog TM Scale-to-PC Data Logging Software OG Version 1.
!!!! CALIBRATION WARNING !!!! Calibration AND inspection of calibration properties is prohibited unless done so by a qualified scale technician. If digital indicator was purchased as part of a complete scale, it is FACTORY CALIBRATED AND READY FOR USE. WARNING The indicator has static sensitive components. Do not make any connections without powering off the indicator. DANGER Electric shock risk. Take necessary precautions to avoid risk of shock. Triner Scale & Mfg.Co., Inc.
Operating Guide | TS-700 Series Table of Contents Specifications TS-700 MS, SS and WB.................................................. 2 Quick Start: Connecting and Powering On TS-700 MS ....................................................................... 3 TS-700 SS ........................................................................ 5 TS-700 WB....................................................................... 7 Operation Keys and Functions: TS-700 MS, SS and WB ..............
Operating Guide | TS-700 Series Specifications Resolution ......................... Display: 30,000 ADC................................... 2,000,000 Zero stability error.............. TK0 < 0.1µV//K Span stability error............. TKspn < ± 6 ppm//K Sensitivity (internal) ........... 0. 3 µV /d Input voltage ...................... -30~30mV DC Excitation circuit................. 5 VDC Load Cell Max.................... 8, 350Ω AC power........................... AC100~250V Operation temp ..................
Operating Guide | TS-700 Series Quick Start: Connecting & Powering On TS-700 MS Connect Weighing Platform Cable Attach the home run cable from the weighing platform to the loadcell port located on the underside of the digital indicator.
Operating Guide | TS-700 Series Quick Start: Connecting & Powering On TS-700 MS (CONT’D) device to the RS-232 port located on the underside of the digital indicator. Power on the device. Connect AC Adapter Plug the AC adapter included with the digital indicator into the 9VDC 1200ma port located on the underside of the digital indicator. Power On The Digital Indicator Press and hold the ON/OFF key for 2+ seconds to power on the indicator. A self-test routine will run, followed by “0” indicated weight.
Operating Guide | TS-700 Series Quick Start: Connecting & Powering On TS-700 SS Connect Weighing Platform Cable Attach the home run cable from the loadcell port to the weighing platform junction box. Refer to installation instructions provided with the Triner Scale weighing platform.
Operating Guide | TS-700 Series Quick Start: Connecting & Powering On TS-700 SS (CONT’D) Connect Any Optional Devices Referring to the manual provided with optional device (such as a receipt printer), connect the device to the RS-232 port located on the underside of the digital indicator. Power on the device. Connect AC Adapter Plug the AC adapter included with the digital indicator into the 9VDC 1200ma port located on the underside of the digital indicator.
Operating Guide | TS-700 Series Quick Start: Connecting & Powering On TS-700 WB Connect the Weigh Bar Cables Attach the home run cables to the loadcell ports and to the weigh bars. For weigh bar set up, refer to the User’s Guide included with livestock the scale.
Operating Guide | TS-700 Series Quick Start: Connecting & Powering On TS-700 WB (CONT’D) Connect Any Optional Devices Referring to the manual provided with optional device (such as a receipt printer), connect the device to the RS-232 port located on the underside of the digital indicator. Power on the device. Connect AC Adapter Plug the AC adapter included with the digital indicator into the 9VDC 1200ma port located on the underside of the digital indicator.
Operating Guide | TS-700 Series Operation KEYS AND FUNCTIONS: TS-700 MS, SS & WB KEY FUNCTION/OPERATION On/Off Press and hold for 2+ seconds to operate. Hold To Hold Weight on Scale (manual hold): With weight on scale, press Hold key. To Hold Peak Weight of Multiple Weighings (auto hold)*: To enable, press Hold key before loading scale. To Hold Weight of an Unstable Load (auto hold)*: To enable, press hold key before loading scale.
Operating Guide | TS-700 Series KEY FUNCTION/OPERATION Total To Add Weights: With weight on scale, press the Total key to place weight and count in memory. Remove weight, place next item on scale, press Total to add weight to total weight and total count. To View Totals: Press Total and Print key at the same time. The total weight will flash continuously. To Print Totals: Press and hold the Print key for 3 seconds while in View Totals mode (requires optional printer).
Operating Guide | TS-700 Series KEY FUNCTION/OPERATION Gross For use after a TARE weight has been stored in the indicator. Press key to toggle view between gross and net weight. PLEASE NOTE that there may be a momentary pause in the display when going from gross to tare weight. In Settings Mode: Right Arrow Key Press to move active digit one unit over to the right Tare With weight on the scale, press to Tare the weight. Press again to remove Tare weight.
Operating Guide | TS-700 Series Parameters and Settings CAUTION: Use care when accessing and adjusting your indicator’s parameters. Several parameters can be accessed that, if altered, will change your indicator’s operations. HOW TO ACCESS AND ADJUST THE TS-700 MS PARAMETERS Press and hold down the PRINT and HOLD keys. “C” will display. Release the keys. “C08” or “C01” will display*, with last digit flashing.
Operating Guide | TS-700 Series Press the kg-lb/left arrow key to move the active digit from the 6 to the 0 (the 0 will flash). Use the Zero/up arrow key to increase the 0 to 1 (C16 will display). Press the Print/enter key. The display will show digits for month, day and year. Use the arrow keys as discussed above to navigate to and change the digits. Press the Print/enter key to accept your settings. Press the Total/home ( ) key to exit.
Operating Guide | TS-700 Series PARAMETERS CHART (For calibration parameters, see Appendix for Scale Technicians, page A-4) PARAMETER FUNCTION OPTIONS C08 Warning tone C09 Automatic Power-off 0 = Disable auto off 10 = Power off in 10 minutes. 30 = Power off in 30 minutes. 60 = Power off in 60 minutes. C10 Power saving setting (display backlight on/off) 0 = Disable power saving setting 3 = Display off in 3 min. 5 = Display off in 5 min.
Operating Guide | TS-700 Series PARAMETER FUNCTION OPTIONS C15 Inner Code display Enter C15 to check the inner code. C16 Date Set the date, from left to right: year/month/day C17 Time Set the time from left to right: hour/min./sec. C18 Communication setting (Serial interface data output method) 0 = Serial interface data output disabled 1 = Continuous sending, remote display 2 = Print key sends data to printer or PC. 3 = PC command request (DataLogPC software) 4 = PC continuous sending.
Operating Guide | TS-700 Series PARAMETER FUNCTION OPTIONS C22 CAUTION: Changing these parameters will effect the indicator’s performance. Automatic Zero: tracking range 0 = Disable zero tracking 0.5 = ±0.5d 1.0 = ±1.0d 2.0 = ±2.0d 3.0 = ±3.0d 4.0 = ±4.0d 5.0 = ±5.0d Note: • d = division • The zero tracking range cannot be more than manual zero range.
Operating Guide | TS-700 Series PARAMETER FUNCTION C26 CAUTION: Changing these parameters will affect the indicator’s performance. Standstill time C27 OPTIONS Standstill range 0 = quick 1 = medium 2 = slow 1 = 1d 2 = 2d 5 = 5d 10 = 10d (d = division) C28 CAUTION: Changing these parameters will affect the indicator’s performance. Digital filter To compensate for unstable loads on the weighing platform (for example, animal weighing).
Operating Guide | TS-700 Series PARAMETER FUNCTION OPTIONS C33 Relay output setting (optional) 0 = Close relay output 1 = Open relay output function 1 2 = Open relay output function 2 3 = Reserved menu C34 RESERVED ------- C35 RESERVED -------- C36 CAUTION: Changing these parameters will affect the indicator’s performance. Gravity of calibration location C36 = 9.7000 -- 9.9999 C37 CAUTION: Changing these parameters will affect the indicator’s performance.
Operating Guide | TS-700 Series Using A Receipt Printer Typical receipt To print receipts with an optional printer: - With power off, attach the printer cable to the connector socket on the underside of the indicator. - Power on the printer & indicator. - Place weight on the scale and press the Print key. - If receipt does not print, settings may need to be adjusted in parameters C18 & C19. - To adjust the format of how the date prints, see options in parameter C30.
Operating Guide | TS-700 Series Appendix For Scale Technicians Calibrating a TS-700 Series Indicator NOTICE: The TS-700 SS/MS/WB are NTEP Class III Legal for Trade approved and certified. Unlocking the calibration parameters is not allowed by persons other than qualified scale technicians. DEFAULT SETTINGS CAUTION Placing the indicator in calibration mode enables access to the Default Settings Parameter, CO7.
Operating Guide | TS-700 Series Appendix For Scale Technicians Calibration Switch, TS-700 MS: Remove the screws and the back plate of the enclosure. A white-capped calibration switch is located at the bottom left corner of the circuit board. Press the switch fully in and release. This will place the switch in the up position, allowing access to the calibration parameters. Loosely reattach the rear plate. Calibration Switch, TS-700 SS: Remove the screws and the back plate of the enclosure.
Operating Guide | TS-700 Series Appendix For Scale Technicians ALL MODELS: After enabling access to calibration parameters, proceed as follows: CALIBRATION PROCEDURE 1) Power on the indicator. 2) Press the PRINT and HOLD keys. “C” will display. Release the keys. 3) “C01” will display, with the 1 flashing. The indicator is now is Parameter Settings mode. Navigating in Settings Mode When the indicator is in settings mode, refer to the graphic printed below the key to determine the keypad key’s function.
Operating Guide | TS-700 Series Appendix For Scale Technicians PARAMETER FUNCTION OPTIONS C01 Default weighing unit 1 = kg 2 = lb Description: With C01 flashing, press the key, then use the up or down arrow keys to change the parameter setting as needed. Press the key to accept the setting and continue on to the next parameter.
Operating Guide | TS-700 Series Appendix For Scale Technicians PARAMETER FUNCTION OPTIONS C03 Graduations 1 = one unit 2 = two units 5 = five units 10 = ten units 20 = twenty units 50 = fifty units Description: With C03 flashing, press the key, then use the up or down arrow keys to change the parameter setting as needed. Press the key to accept the setting and continue on to the next parameter.
Operating Guide | TS-700 Series Appendix For Scale Technicians PARAMETER FUNCTION OPTIONS C05 Zero Calibration 0 = skip zero calibration 1 = proceed with zero calibration Description: With C05 flashing, press the key, then use the up arrow key to change the 0 parameter setting to 1. Press the key to proceed. The zero cal will count down from 10. Press the key to proceed to the next parameter.
Operating Guide | TS-700 Series Appendix For Scale Technicians PARAMETER FUNCTION C06 Calibration OPTIONS 0 = skip calibration 1 = proceed with calibration Description: With C06 flashing, press the key, then use the up arrow key to change the parameter setting to 1. Press the key to proceed. “Span” will briefly display, then all digits will display. Load the weighing platform, then adjust the displayed digits to match the weight on the platform, i.e, 4,000. Press the key to proceed.
Operating Guide | TS-700 Series Appendix For Scale Technicians PARAMETER FUNCTION C07 CAUTION!! FACTORY RESET… Erases all custom settings! OPTIONS 0 = skip reset to factory defaults 1 = proceed with reset to defaults Description: To reset to factory defaults, change the 0 setting to 1 and press the key.
Operating Guide | TS-700 Series Appendix For Scale Technicians Load Cell Connections The indicator can connect with a maximum of eight (8) 350Ω load cells, 4 wire or 6 wire configurations. Excitation voltage for the load cell is 5VDC, the maximum output current is 120mA.
Operating Guide | TS-700 Series Appendix For Scale Technicians Serial Port Connection RS232 CONNECTION TS-700 MS The RS232 DB9 serial port can be utilized to connect the indicator to a peripheral device such as a printer or remote display, or to interface the indicator with a PC. NOTE: PC interface requires Triner DataLog PC™ software to be installed on the PC.
Operating Guide | TS-700 Series Appendix For Scale Technicians TS-700 SS The 3-wire terminal block located on the main board can be utilized to connect the indicator to a peripheral device such as a printer or remote display, or to interface the indicator with a PC. NOTE: PC interface requires Triner DataLog PC™ software to be installed on the PC.
Operating Guide | TS-700 Series Appendix For Scale Technicians Data Output Formats REMOTE DISPLAY FORMAT State A Bits: 0,1,2 0 1 2 1 0 0 0 1 0 1 1 1 0 0 Bits: 3,4 0 0 1 1 0 1 Decimal point position XXXXXX0 XXXXXXX XXXXX-X XXXX-XX XXX-XXX Division X1 X2 1 0 State B BITS FUNCTION Bits0 Gross = 0, net=1 Bits1 Symbol: positive = 0,negative =1 Bits2 Overload (or under zero)=1 Bits3 Dynamic = 1 Bits4 Unit: lb=0, kg=1 Bits5 Bits6 Constant 1 Constant 0 State C Bit2 0 0 0 Bit1 0 0 1 Bit 3
Operating Guide | TS-700 Series Appendix For Scale Technicians PC COMPUTER CONTINUOUS SENDING FORMAT , S1 , S2 CR S3 Data S1: Weight status: ST = standstill US = not standstill OL = overload S2: Weight mode: GS = gross mode NT = net mode S3: Weight of positive and negative, “+” or ” –“ S4: “kg” or “lb” Data: Weight value, including decimal point CR: Carriage return LF: Line feed A-13 S4 L F
Operating Guide | TS-700 Series Appendix For Scale Technicians SERIAL INTERFACE RECEPTION COMMAND RS232COM serial interface can receive simple ASCII command. Command word and role as follows: Command T Z P R Name TARE ZERO PRINT G.W/N.W Function Save and clear tare Zero gross weight Print the weight Read gross weight or net weight C Kg/lb Kg/lb conversion G G.W Check gross weight at net weight mode Data Format Print Format ID.NO. ....................004 (Serial No.) Date:.......................
Operating Guide | TS-700 Series Appendix For Scale Technicians Error Codes ERROR CAUSE SOLUTION 1. Overloaded condition. 2. Wrong connection at load cell. 3. Load cell has quality problem. 1. Reduce the weight. 2. Check load cell connection 3. Inspect load cell. Check the input and output. 1. Calibration is incorrect. 2. Bad connections. 3. Load cell has quality problem. 1. Check scale is resisted or not, foot is kept level or not. 2. Check load cell connection. 3.
Operating Guide | TS-700 Series Appendix For Scale Technicians ERROR CAUSE SOLUTION ERR4 During calibration, the signal is unstable. After the platform is stable, start calibration. ERR5 Circuitry error Replace the PCB. Factory Default Settings PARAMETER INSTRUCTION DEFAULT C01 Calibration 1 C02 Decimal digits 0 C03 Resolution 1 C04 Max.
Operating Guide | TS-700 Series Appendix For Scale Technicians PARAMETER INSTRUCTION DEFAULT C13 Upper limit alarm 000000 C14 Under limit alarm 000000 C15 Inner code ---- C16 Date setting ---- C17 Time setting ---- C18 Serial interface data output C19 Serial interface Baud rate C20 Zero manually 10 C21 Initial zero 10 C22 Zero tracking range 0 C23 Zero tracking time 1 C24 Overload range 9 C25 Negative range 10 C26 Standstill time 1 C27 Standstill range 2 C28 Dyn
Operating Guide | TS-700 Series Appendix For Scale Technicians PARAMETER INSTRUCTION DEFAULT C30 Print format 0 C31 Analog signal options 1 C32 4~20mA testing 4 C33 Relay output setting 1 C34 Reserved 0 C35 Reserved 6 C36 Calibration location gravity 9.7936 C37 Destination gravity 9.7936 C38 Version No.