User guide
Typical Applications Include
! Semiconductor
! LCD
! Electronic assembly
! ESD-sensitive processes
The Trek Model 541A Electrostatic Voltmeter (pictured at left with
the Model 541PR-S probe) provides accurate non-contacting
measurements of the electrostatic surface voltage associated
with EOS/ESD processes. The instrument is configured with a
miniature electrostatic field chopper probe that can be remotely
located and easily positioned within process equipment to provide
highly accurate, non-contacting, DC-stable, spacing-independent
voltage measurements in either ionized or non-ionized
environments. A 20x4 alphanumeric LCD screen displays the
present measured voltage, the positive peak voltage, the negative
peak voltage and additional menu information.
Trek Model 541A Non-Contacting
Electrostatic Voltmeter for EOS/ESD
Key Specifications
!
!
!
Measurement Range
Model 541A-1: ±1 kV DC or peak AC
Model 541A-2: ±100 V DC or peak AC
Measurement Accuracy: Better than ±1% of full scale over a
probe-to-surface separation of
2.5 mm ± 1mm
Alphanumeric LCD Display: 20 x 4 characters
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
716-438-7555 • 716-201-1804 (fax) • www.trekinc.com • sales@trekinc.com
Features and Benefits
! Chopper probe is DC-stable with or without incident air ion flow
! Drift-free measurements
! LCD screen displays present voltage and holds the most positive and negative values
! Visual and audible alarms activate when the preset voltage threshold levels are reached
! Voltage output monitor for remote monitoring or control
! USB and RS-232 serial ports
! NIST-traceable Certificate of Calibration provided with each unit
! compliant
! Optional Walking Test Adapter kits available
Walking Test Adapter
Optional Walking Test Adapter
kits for the Model 541A allow
analysis of charge levels
accumulated on the human body
per compliance with ANSI/ESD
STM97.2. and IEC Standard
61340-4-5. See page 2 for details.