Users Manual
Appendix
Table 3
Guidance and manufacturer’s declaration - electromagnetic Immunity
Radiated RF
IEC61000-4-3
(Test specifications
for ENCLOSURE
PORT IMMUNITY to
RF wireless
communications
equipment)
Test
Frequency
(MHz)
Band
(MHz)
Service Modulation Modulation
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
385 380 –
390
TETRA 400 Pulse
modulation
18 Hz
1,8 0.3 27
450 430 –
470
GMRS 460,
FRS 460
FM
±5kHz
deviation
1kHzsine
20.328
710 704 –
787
LTE Band
13,
17
Pulse
modulation
217 Hz
0,2 0.3 9
745
780
810 800 –
960
GSM
800/900,
TETRA
800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
20.328
870
930
1720 1 700 –
1990
GSM 1800;
CDMA
1900;
GSM 1900;
DECT;
LTE Band
1, 3,
4, 25;
UMTS
Pulse
modulation
217 Hz
20.328
1845
1970
2450 2 400 –
2570
Bluetooth,
WLAN,
802.11
b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
20.328