Data Sheet

Safety Standards
- IT / Multimedia Equipment
EN 62368-1
IEC 62368-1
UL 62368-1
- Railway Fire Protection
EN 45545-2
- Medical Equipment
EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
- Certification Documents
www.tracopower.com/overview/tpp450ba-m
Protection Class
Class I & II (Prepared): Reinforced Insulation
Pollution Degree
PD 2
Over Voltage Category
OVC II
Safety Specifications
EMI Emissions
EN 60601-1-2 edition 4 (Medical Devices)
- Conducted Emissions
EN 55011 class B (internal filter)
EN 55032 class B (internal filter)
FCC Part 15 class B (internal filter)
FCC Part 18 class B (internal filter)
- Radiated Emissions
EN 55011 class A (internal filter)
EN 55032 class A (internal filter)
FCC Part 15 class A (internal filter)
FCC Part 18 class A (internal filter)
- Harmonic Current Emissions
EN 61000-3-2, class A
EN 61000-3-2, class D
- Voltage Fluctuations & Flicker
EN 61000-3-3
(For optimal EMI performance the power supply
should be mounted to a grounded aluminium
plate (480 x 248 x 12 mm) with electrical contact
to the four PCB mounting holes. To comply with
safety standards, this plate must be grounded.)
EMS Immunity
EN 55024 (IT Equipment)
EN 60601-1-2 edition 4 (Medical Devices)
- Electrostatic Discharge
Air:
EN 61000-4-2, ±15 kV, perf. criteria A
Contact:
EN 61000-4-2, ±8 kV, perf. criteria A
- RF Electromagnetic Field
EN 61000-4-3, 3 V/m, perf. criteria A
- EFT (Burst) / Surge
EN 61000-4-4, ±2 kV, perf. criteria A
L to L:
EN 61000-4-5, ±1 kV, perf. criteria A
L to PE:
EN 61000-4-5, ±2 kV, perf. criteria A
- Conducted RF Disturbances
EN 61000-4-6, 20 Vrms, perf. criteria A
- PF Magnetic Field
Continuous:
EN 61000-4-8, 30 A/m, perf. criteria A
- Voltage Dips & Interruptions
230 VAC / 50 Hz:
EN 61000-4-11
30%, 25 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria B
115 VAC / 60 Hz:
EN 61000-4-11
30%, 25 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria B
EMC Specifications
Relative Humidity
95% max. (non condensing)
Temperature Ranges
- Operating Temperature
–40°C to +85°C
- Storage Temperature
–40°C to +85°C
General Specifications
TPP 450BA-M Series, 450 Watt
All specifications valid at nominal voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com
March 29, 2021
Page 3 / 5