Datasheet

EMI Emissions
EN 60601-1-2 edition 4 (Medical Devices)
EN 55011 class B (internal filter)
EN 55032 class B (internal filter)
FCC Part 15 class B (internal filter)
FCC Part 18 class B (internal filter)
EN 55011 class B (internal filter)
EN 55032 class B (internal filter)
FCC Part 15 class B (internal filter)
FCC Part 18 class B (internal filter)
EN 61000-3-2, class A
EN 61000-3-3
EMS Immunity
EN 60601-1-2 edition 4 (Medical Devices)
Air:
EN 61000-4-2, ±15 kV, perf. criteria A
Contact:
EN 61000-4-2, ±8 kV, perf. criteria A
EN 61000-4-3, 20 V/m, perf. criteria A
EN 61000-4-4, ±2 kV, perf. criteria A
L to L:
EN 61000-4-5, ±1 kV, perf. criteria A
L to PE:
EN 61000-4-5, ±2 kV, perf. criteria A
EN 61000-4-6, 20 Vrms, perf. criteria A
Continuous:
EN 61000-4-8, 30 A/m, perf. criteria A
230 VAC / 50 Hz:
EN 61000-4-11
30%, 25 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria B
115 VAC / 60 Hz:
EN 61000-4-11
30%, 25 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria B
EMC Specifications
Relative Humidity
95% max. (non condensing)
Temperature Ranges
–40°C to +85°C
–40°C to +85°C
Power Derating
Depending on model
Depending on model
See application note:
www.tracopower.com/overview/tpp40a
Cooling System
Natural convection (20 LFM)
Altitude During Operation
5'000 m max.
Switching Frequency
50 - 140 kHz (PWM)
Insulation System
Reinforced Insulation
Working Voltage (rated)
258 VAC
Isolation Test Voltage
4'000 VAC
2'500 VAC
2'500 VAC
Creepage
8 mm min.
Clearance
8 mm min.
Isolation Resistance
100 min.
Leakage Current
(at 264 VAC)
75 µA max.
Reliability
3'000'000 h (MIL-HDBK-217F, ground benign)
General Specifications
TPP 40A Series, 40 Watt
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com September 20, 2023
Page 3 / 4