AC/DC Medical Power Supply TPP 30-D Series, 30 Watt
•
High power density power supply
(encapsulated)
•
Certification according to IEC/EN/ES
60601-1 3rd edition for 2 x MOPP
•
Low leakage current <75 µA rated for
BF applications
•
EMC compliance to IEC 60601-1-2 4th
edition
•
Risk management process according to
ISO 14971 incl. risk management file
•
Acceptance criteria for electronic
assemblies acc. to IPC-A-610 Level 3
•
Protection class ll
•
Operating up to 5000 m altitude
•
Ready to meet ErP directive, no load
power consumption <60 mW
•
5-year product warranty
IEC 60601-1
IEC 60335-1
IEC 62368-1
ES 60601-1
UL 62368-1
The TPP 30-D AC/DC power supplies feature a reinforced double I/O isolation
system according to medical safety standards IEC/EN/ES 60601-1 3rd edition
for 2 x MOPP approved for an operating altitude of 5000 m. The earth leakage
current is below 75 µA what makes the units suitable for BF (body floating) ap-
plications. The excellent efficiency of up to 92% offers a high power density in
the packaging format 1.5” x 2.9”. The full load operating temperature range
covers –40°C to +60°C while it goes up to 85°C with 50% load derating. The
units operate in compliance to the medical EMC emission and immunity levels
according to latest standard IEC 60601-1-2 4th edition.
Models
Order Code
Output Power
Output Voltage
Output Current
Efficiency
max.
nom. (adjustable)
max.
typ.
TPP 30-103-D
20 W
3.3 VDC (2.97 - 3.63 VDC)
6'000 mA
84 %
TPP 30-105-D
30 W
5 VDC (4.5 - 5.5 VDC)
6'000 mA
87 %
TPP 30-109-D
9 VDC (8.1 - 9.9 VDC)
3'340 mA
88 %
TPP 30-112-D
12 VDC (10.8 - 13.2 VDC)
2'500 mA
91 %
TPP 30-115-D
15 VDC (13.5 - 16.5 VDC)
2'000 mA
91 %
TPP 30-124-D
24 VDC (21.6 - 26.4 VDC)
1'250 mA
90 %
TPP 30-136-D
36 VDC (32.4 - 39.6 VDC)
840 mA
90 %
TPP 30-148-D
48 VDC (43.2 - 52.8 VDC)
630 mA
92 %
www.tracopower.com September 20, 2023