Data Sheet
ht tp://www.tracopower.com Page 1 of 4
DC/DC Converter   THM 10 Series, 10 Watt
The THM-10 series is a range of medical 10 Watt DC/DC converters in DIP-24 
plastic package and with wide 2:1 input voltage range. They provide a reinforced 
isolation system for 5000 VACrms isolation and a very low leakage current of 
less than 2 A. The units are approved to IEC/EN/ES 60601-1 3rd edition for 2 
x MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk 
management file. Design and production conform to the quality management 
system ISO 13485. With a high efficiency of up to 89% and highest grade com-
ponents the converters can reliably operate in an ambient temperature range 
of –40°C up to +90°C. They constitute a reliable solution not only for medical 
equipment but also for demanding ranges of application such as transportation, 
control & measurement or IGBT drivers.
•  Wide 2:1 input voltage
•  I/O isolation 5000 VACrms rated for 
  250 VACrms working voltage 
•  Certification according to IEC/EN/ES   
  60601-1 3rd edition for 2MOPP
•  Risk management process according to 
  ISO 14971 including risk management file
•  Acceptance criteria for electronic 
  assemblies according to IPC-A-610 Level 3
•  Low leakage current
•  Extended operating temperature range  
  –40°C to 90°C.
•  Input filter to meet EN55022 class A
•  Operating up to 5000m altitude
•  5 year product warranty
Order code Input voltage range Output voltage Output current max. Efficiency typ.
THM 10-0510
  3.3 VDC   2500 mA 80.0 %
THM 10-0511
  5.0 VDC   2000 mA 84.0 %
THM 10-0512
  12 VDC   830 mA 86.5 %
THM 10-0513
4.5 – 9 VDC
  15 VDC   670 mA 87.0 %
THM 10-0515 (5 VDC nominal)
  24 VDC   416 mA 85.5 %
THM 10-0521
  ±5.0 VDC   ±1000 mA 83.0 %
THM 10-0522
  ±12 VDC   ±416 mA 85.5 %
THM 10-0523
  ±15 VDC   ±333 mA 86.5 %
THM 10-1210
  3.3 VDC   2500 mA 83.0 %
THM 10-1211
  5.0 VDC   2000 mA 85.5 %
THM 10-1212
  12 VDC   830 mA 88.0 %
THM 10-1213
9. 0 – 18 VDC
  15 VDC   670 mA 89.0 %
THM 10-1215 (12 VDC nominal)
  24 VDC   416 mA 89.0 %
THM 10-1221
  ±5.0 VDC   ±1000 mA 84.0 %
THM 10-1222
  ±12 VDC   ±416 mA 89.0 %
THM 10-1223
  ±15 VDC   ±333 mA 88.0 %
THM 10-2410
  3.3 VDC   2500 mA 83.0 %
THM 10-2411
  5.0 VDC   2000 mA 86.5 %
THM 10-2412
  12 VDC   830 mA 89.0 %
THM 10-2413
18 – 36 VDC
  15 VDC   670 mA 89.0 %
THM 10-2415 (24 VDC nominal)
  24 VDC   416 mA 89.0 %
THM 10-2421
  ±5.0 VDC   ±1000 mA 85.0 %
THM 10-2422
  ±12 VDC   ±416 mA 89.0 %
THM 10-2423
  ±15 VDC   ±333 mA 88.0 %
THM 10-4810
  3.3 VDC   2500 mA 82.5 %
THM 10 - 4811
  5.0 VDC   2000 mA 86.5 %
THM 10-4812
  12 VDC   830 mA 89.0 %
THM 10-4813
36 – 75 VDC
  15 VDC   670 mA 89.0 %
THM 10-4815 (48 VDC nominal)
  24 VDC   416 mA 88.5 %
THM 10-4821
  ±5.0 VDC   ±1000 mA 85.0 %
THM 10-4822
  ±12 VDC   ±416 mA 88.0 %
THM 10-4823
  ±15 VDC   ±333 mA 88.0 %
Models
ES 60601-1
C B
Scheme
IEC 606010-1