Datasheet

EMC Emission
EN 60601-1-2 edition 4 (Medical Devices)
- Conducted Emissions
EN 55011 class B (internal filter)
EN 55032 class B (internal filter)
FCC Part 15, class B
FCC Part 18, class B
- Radiated Emissions
EN 55011 class B (internal filter)
EN 55032 class B (internal filter)
FCC Part 15, class B
FCC Part 18, class B
- Harmonic Current Emissions
EN 61000-3-2, class A
- Voltage Fluctuations & Flicker
EN 61000-3-3
EMC Immunity
EN 60601-1-2 edition 4 (Medical Devices)
- Electrostatic Discharge
Air:
EN 61000-4-2, ±15 kV, perf. criteria A
Contact:
EN 61000-4-2, ±8 kV, perf. criteria A
- RF Electromagnetic Field
EN 61000-4-3, 20 V/m, perf. criteria A
- EFT (Burst)
EN 61000-4-4, ±2 kV, perf. criteria A
- Surge
L to L:
EN 61000-4-5, ±1 kV, perf. criteria A
L to PE:
EN 61000-4-5, ±2 kV, perf. criteria A
- Conducted RF Disturbances
EN 61000-4-6, 20 Vrms, perf. criteria A
- PF Magnetic Field
EN 61000-4-8, 30 A/m, perf. criteria A
- Voltage Dips & Interruptions
230 VAC / 50 Hz:
EN 61000-4-11
30%, perf. criteria A, 25 periods
>95%, perf. criteria A, 0.5 periods
>95%, perf. criteria A, 1 period
>95%, perf. criteria B, 250 periods
115 VAC / 60 Hz:
EN 61000-4-11
30%, 25 periods, perf. criteria A
>95%, 0.5 periods, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria B
Relative Humidity
95% max. (non condensing)
Temperature Ranges
- Operating Temperature
-40°C to +85°C
- Storage Temperature
-40°C to +85°C
Power Derating
- High Temperature
see application note
www.tracopower.com/overview/tpp40a
Cooling system
Natural convection (no internal fan, 20 LFM)
Altitude during Operation
5000 m max.
Switching Frequency
50 - 140 kHz (PWM)
Insulation System
Reinfoced Insulation
Working Voltage (rated)
258 VAC
Isolation Test Voltage
- Input to Output, 60 s
5'657 VDC
- Input to Case or PE, 60 s
3'535 VDC
- Output to Case or PE, 60 s
3'535 VDC
Isolation Resistance
- Input to Output, 500 VDC
100 MOhm min.
Leakage Current
(at 264 VAC)
- Touch Current
75 µA max.
Creepage
8 mm min.
Clearance
8 mm min.
Reliability
- Calculated MTBF
3'000'000 h (acc. to MIL-HDBK-217F)
Environment
- Vibration
IEC 60068-2-6
3 axis, sine sweep, 10 - 55 Hz, 1 g, 1 oct/min
- Mechanical Shock
IEC 60068-2-27
3 axis, 10 g half sine, 11 ms shock
20 g (3 directions each 3 times)
EMC Specifications
General Specifications
TPP 40A Series, 40 Watt
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com
March 18, 2019
Page 3 / 4