Datasheet

www.tracopower.com Page 3 of 5
TMF Series, 5 – 30 W
General Specifications (continued)
Protection class class II prepared
Safety standards IEC 60601-1:2005 (3rd Ed.)+ A1:2012,
ANSI/A AMI ES60601-1:2005/(R)2012,
CSA CAN/CSA-C22.2 NO. 60601-1:14
EN 60601-1:2006 + A1:2013 + A12:2014
– Certification documents www.tracopower.com/overview/tmf
Electromagnetic compatibility (EMC), Emissions
– Conducted & Radiated input suppression EN 55011 limits to IEC 60601-1-2 4th editon
Electromagnetic compatibility (EMC), Immunity
– Electrostatic discharge (ESD) IEC/EN 61000-4-2 8 kV / 15 kV criteria A
– Radiated RF field immunity IEC/EN 61000-4-3 10 V/m criteria A
– Electrical fast transient / burst immunity IEC/EN 61000-4-4 2 kV criteria A
– Surge immunity* IEC/EN 61000-4-5 1 kV criteria A
– Immunity to conducted RF disturbances IEC/EN 61000-4-6 10 V criteria A
– Magnetic field immunity IEC/EN 61000-4-8 30 A/m criteria A
– Mains voltage dips and interruptions IEC/EN 61000-4-11
30% / 10ms criteria B
95% / 5000ms criteria B
Case material plastic (UL94V-0 rated)
Environmental compliance – Reach www.tracopower.com/info/reach-declaration.pdf
– RoHS RoHS directive 2011/65/EU
* For the 5 & 10 W models an external Varistor has to be integreated in the circuit at the converter input (in parallel).
Varistor recommendation: 14S471K