Datasheet
DC/DC Medical Converter
THM 10 Series, 10 Watt
•
Compact DIP-24 plastic case
•
I/O isolation 5000 VAC rated for 250
VAC working voltage
•
IEC 60601-1 certification for 2 x MOPP
•
Risk management process according to
ISO 14971 incl. risk management file
•
Acceptance criteria for electronic
assemblies acc. to IPC-A-610 Level 3
•
Low leakage current <2 µA
•
Operating temperature –40°C to 90°C
•
EMC compliance to IEC 60601-1-2 4th
edition and EN55032 class A
•
5-year product warranty
IEC 60601-1
ES 60601-1
The THM 10 series is a range of medical 10 Watt DC/DC converters in DIP-24
plastic package with wide 2:1 input voltage range. They provide a reinforced
isolation system for 5000 VAC and a very low leakage current of less than 2 µA.
The units are approved to IEC/EN/ES 60601-1 3rd ed. for 2 x MOPP and
come along with an ISO 14971 risk management file. Design and production
conform to the quality management system ISO 13485. With a high efficiency
of up to 89% and highest grade components the converters can reliably oper-
ate in an ambient temperature range of –40°C up to +90°C. They constitute a
reliable solution not only for medical equipment but also for demanding ranges
of application such as transportation, control & measurement or IGBT drivers.
Models
Order Code
Input Voltage
Output 1
Output 2
Efficiency
Range
Vnom
Imax
Vnom
Imax
typ.
THM 10-0510
4.5 - 9 VDC
(5 VDC nom.)
3.3 VDC
2'500 mA
80 %
THM 10-0511
5 VDC
2'000 mA
84 %
THM 10-0512
12 VDC
830 mA
87 %
THM 10-0513
15 VDC
670 mA
87 %
THM 10-0515
24 VDC
416 mA
86 %
THM 10-0521
+5 VDC
1'000 mA
–5 VDC
1'000 mA
83 %
THM 10-0522
+12 VDC
416 mA
–12 VDC
416 mA
86 %
THM 10-0523
+15 VDC
333 mA
–15 VDC
333 mA
87 %
THM 10-1210
9 - 18 VDC
(12 VDC nom.)
3.3 VDC
2'500 mA
83 %
THM 10-1211
5 VDC
2'000 mA
86 %
THM 10-1212
12 VDC
830 mA
88 %
THM 10-1213
15 VDC
670 mA
89 %
THM 10-1215
24 VDC
416 mA
89 %
THM 10-1221
+5 VDC
1'000 mA
–5 VDC
1'000 mA
84 %
THM 10-1222
+12 VDC
416 mA
–12 VDC
416 mA
89 %
THM 10-1223
+15 VDC
333 mA
–15 VDC
333 mA
88 %
THM 10-2410
18 - 36 VDC
(24 VDC nom.)
3.3 VDC
2'500 mA
83 %
THM 10-2411
5 VDC
2'000 mA
87 %
THM 10-2412
12 VDC
830 mA
89 %
THM 10-2413
15 VDC
670 mA
89 %
THM 10-2415
24 VDC
416 mA
89 %
THM 10-2421
+5 VDC
1'000 mA
–5 VDC
1'000 mA
85 %
THM 10-2422
+12 VDC
416 mA
–12 VDC
416 mA
89 %
THM 10-2423
+15 VDC
333 mA
–15 VDC
333 mA
88 %
THM 10-4810
36 - 75 VDC
(48 VDC nom.)
3.3 VDC
2'500 mA
83 %
THM 10-4811
5 VDC
2'000 mA
87 %
THM 10-4812
12 VDC
830 mA
89 %
THM 10-4813
15 VDC
670 mA
89 %
THM 10-4815
24 VDC
416 mA
89 %
THM 10-4821
+5 VDC
1'000 mA
–5 VDC
1'000 mA
85 %
THM 10-4822
+12 VDC
416 mA
–12 VDC
416 mA
88 %
THM 10-4823
+15 VDC
333 mA
–15 VDC
333 mA
88 %
www.tracopower.com
September 16, 2019
Page 1 / 5