Datasheet
Output Current Limitation
150% typ. of Iout max.
Overvoltage Protection
112 - 152% of Vout nom.
(depending on model)
3.7 - 5 VDC (3.3 VDC model)
5.6 - 7 VDC (5 VDC model)
13.5 - 16 VDC (12 VDC model)
18.3 - 22 VDC (15 VDC model)
29.1 - 34.5 VDC (24 VDC model)
5.6 - 7 VDC (±5 VDC model)
13.5 - 18.2 VDC (±12 VDC model)
17 - 22 VDC (±15 VDC model)
Transient Response
- Response Time
250 µs typ. (25% Load Step)
Safety Standards
- Medical Equipment
EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
- Certification Documents
www.tracopower.com/overview/thm10wi
Pollution Degree
PD 2
Over Voltage Category
OVC II
Safety Specifications
EMI Emissions
EN 60601-1-2 edition 4 (Medical Devices)
- Conducted Emissions
EN 55011 class A (internal filter)
EN 55011 class B (with external filter)
EN 55032 class A (internal filter)
EN 55032 class B (with external filter)
FCC Part 18, class A
FCC Part 18, class B
- Radiated Emissions
EN 55011 class A (internal filter)
EN 55011 class B (with external filter)
EN 55032 class A (internal filter)
EN 55032 class B (with external filter)
FCC Part 18, class A
FCC Part 18, class B
External filter proposal:
www.tracopower.com/overview/thm10wi
EMS Immunity
EN 60601-1-2 edition 4 (Medical Devices)
- Electrostatic Discharge
Air:
EN 61000-4-2, ±15 kV, perf. criteria A
Contact:
EN 61000-4-2, ±8 kV, perf. criteria A
- RF Electromagnetic Field
EN 61000-4-3, 10 V/m, perf. criteria A
- EFT (Burst) / Surge
EN 61000-4-4, ±2 kV, perf. criteria A
EN 61000-4-5, ±2 kV, perf. criteria A
Ext. input component:
5 Vin models: KY 1000 µF // Vishay V10P45
24 Vin models: KY 470 µF
48 Vin models: KY 330 µF
- Conducted RF Disturbances
EN 61000-4-6, 10 Vrms, perf. criteria A
- PF Magnetic Field
Continuous:
EN 61000-4-8, 100 A/m, perf. criteria A
1 s:
EN 61000-4-8, 1000 A/m, perf. criteria A
EMC Specifications
Relative Humidity
95% max. (non condensing)
Temperature Ranges
- Operating Temperature
–40°C to +90°C
- Approved Ambient Temp.
+50°C max. (to comply with EN 60601-1)
- Case Temperature
+105°C max.
- Storage Temperature
–55°C to +125°C
Power Derating
- High Temperature
3.33 %/K above 75°C
General Specifications
THM 10WI Series, 10 Watt
All specifications valid at nominal voltage, full load and +25°C after warm-up time unless otherwise stated.
www.tracopower.com
September 16, 2019