Model
AGCN100 AGCN200
Maximum Capacity 50g 100g
Readability (d) 0.001g
Repeatability (Standard
Deviation)
0.001g
Linearity +/- 0.002g
Tare Range -50g -100g
Accuracy Class II
Calibration Weight 50g 100g
Operating Temperature +18C to +33C
Analyzer Resolution 0.01%
Moisture Measurement
Accuracy
0.1g to 5g = +/-0.3%
5g to 15g = +/-0.06%
> 15g = < +/-0.04%
Maximum Drying
Temperature
160°C
Sampling Time 1 sec. to 180 sec.
Maximum Drying Time 10 hours
Drying Power 250W
Halogen Size 118mm
Warm-up time Approximately 30 sec.
Pan Size 90mm
Drying Chamber
Dimensions
108mm x 20mm
Scale Dimensions 215mm x 345mm x 200mm
Unit Weight 7kg
Power 110VAC, 60Hz
Warranty 12 months