User's Manual
74
Tobii C-Series User’s Manual
Chapter 8 Appendices
Guidance and manufacturer’s declaration – electromagnetic immunity
The C-Series device is intended for use in the electromagnetic environment specied below. The
customer of the user of the C-Series device should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the C-Series device, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Pd
3
5.3
=
for 150 kHz to 80 MHz
Pd
3
5.3
=
for 80 MHz to 800 MHz
Pd
3
7
=
for 800 MHz to 2.5 GHz
where “P” is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer and
“d” is the recommended separation distance
in meters (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol.
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reected from structures, objects and people.