User's Manual

Table Of Contents
Guidance and manufacturers declaration electromagnetic emissions
The Tobii Dynavox I-Series+ is intended for use in the electromagnetic environment specified below. The customer or the
user of the Tobii Dynavox I-Series+ device should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment
guidance
RF (Radio Frequency
emissions CISPR 11
Group 1 The Tobii Dynavox I-Series+ must emit
electromagnetic energy in order to
perform its intended function. Nearby
electronic equipment may be affected
RF emissions CISPR 11 Class B
Harmonic emissions
IEC 61000-3-2
Not Applicable The product has an active power of
P75 W at rated load conditions.
Therefore no limits are applicable.
Voltage Fluctuations/flicker
emissions IEC 61000-3-3
Not Applicable The product has an active power of
P75 W at rated load conditions.
Therefore it is not likely to produce
voltage fluctuations or flicker above the
limit values.
The Tobii Dynavox I-Series+ device should not be used adjacent to or stacked with other equipment.
Guidance and manufacturers declaration electromagnetic immunity
The Tobii Dynavox I-Series+ device is intended for use in the electromagnetic environment specified below. The customer or
the user of the Tobii Dynavox I-Series+ device should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 test level Compliance level
Electromagnetic environment
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line to line ± 1 kV line to line Mains power quality should
be that of a typical
commercial or hospital
environment.
80
Appendix B Compliance Information
Tobii Dynavox I-Series+ Users manual v.1.0 - en-US