User's Manual
Table Of Contents
- 1 Introduction
- 2 Safety
- 3 Overview of the Tobii Dynavox I–12+ and I–15+
- 4 Batteries in the Tobii Dynavox I-Series+
- 5 Using the Tobii Dynavox I-Series+
- 5.1 Starting the Device
- 5.2 Shutting Down the Device
- 5.3 Using Gaze Interaction
- 5.4 Using Windows Control
- 5.5 Wake the Tobii Dynavox I-Series+ from Sleep with Gaze Interaction
- 5.6 Pause/Resume Gaze Interaction and Put the Tobii Dynavox I-Series+ to Sleep with Gaze Interaction
- 5.7 Using the Cameras
- 5.8 Using the GEWA Programmable IR Remote Control
- 5.9 Adjusting the Volume
- 5.10 Adjusting the Sound
- 5.11 Performing a Factory Recovery
- 5.12 Calibrating the Touch Screen
- 6 Configuring the Tobii Dynavox I-Series+ with the I-Series Control Center
- 7 Gaze Interaction Settings
- 8 Gaze Selection Settings
- 9 Product Care
- Appendix A Support, Warranty, Training Resources
- Appendix B Compliance Information
- Appendix C Computer Device States of Operation
- Appendix D Maximum Allowed Temperature
- Appendix E Technical Specifications
- Appendix F Approved Accessories
Guidance and manufacturer’s declaration – electromagnetic emissions
The Tobii Dynavox I-Series+ is intended for use in the electromagnetic environment specified below. The customer or the
user of the Tobii Dynavox I-Series+ device should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment –
guidance
RF (Radio Frequency
emissions CISPR 11
Group 1 The Tobii Dynavox I-Series+ must emit
electromagnetic energy in order to
perform its intended function. Nearby
electronic equipment may be affected
RF emissions CISPR 11 Class B
Harmonic emissions
IEC 61000-3-2
Not Applicable The product has an active power of
P≤75 W at rated load conditions.
Therefore no limits are applicable.
Voltage Fluctuations/flicker
emissions IEC 61000-3-3
Not Applicable The product has an active power of
P≤75 W at rated load conditions.
Therefore it is not likely to produce
voltage fluctuations or flicker above the
limit values.
The Tobii Dynavox I-Series+ device should not be used adjacent to or stacked with other equipment.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Tobii Dynavox I-Series+ device is intended for use in the electromagnetic environment specified below. The customer or
the user of the Tobii Dynavox I-Series+ device should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 test level Compliance level
Electromagnetic environment
– guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line to line ± 1 kV line to line Mains power quality should
be that of a typical
commercial or hospital
environment.
80
Appendix B Compliance Information
Tobii Dynavox I-Series+ User’s manual v.1.0 - en-US