User's Manual
Customer/Project:
Sotec
Reference:
Rev: B
User Guide of TH30 and TRF2
Release Date: 04/01/2017
Pages 5 / 6
4. Un-pairing procedure
A. Please put TH30 and TRF2 together with the distance less than 50cm like this picture.
B. Kee
p pressing button 3 and button 4 for 7 seconds to activate un-pairing mode, un-pairing procedure completed after LED
indicator flashing 3 times
A. Please put TH30 and TRF2 together with the distance less than
50cm like this picture.
B. Kee
p pressing button 3 and button 4 for 7 seconds to activate
un-pairing mode, un-pairing procedure completed after LED
indicator flashing 3 times
Remark:
Note 1: Un-pairing procedure only done by TH30.
5. Product Application:
TH30 is for hospital, home beds and nursing beds. The operators will use the wireless handset to operate bed frame, and
avoid contacting with the handset which the patient used.
6. Others
1.
Modulation: GFSK
2.
Operating frequency: 2420MHz, 2450MHz, 2480MHz
3. Power supply: TH30 (AA battery) / TRF2 (Timotion RJ plug)
4. Input voltage : TH30 (DC 3V) / TRF2 (DC 24〜36V)
5. Environmental Requirements
Operation conditions
Ambient temperature: 10 ~ 40℃ ℃
Relative humidity range: 30% ~ 75%
Atmospheric pressure: 86KPa~ 106Kpa
Storage and shipping conditions
Ambient temperature: -10 ~ ℃ 50℃
Relative humidity range: 10% ~ 93%
Atmospheric pressure: 86KPa~ 106KPa