2.6 Safety Notes
Danger/Warning:
• Keep this device out of the reach of children.
• Do not immerse the device in liquids.
• Do not use while driving motor vehicles or operating heavy
machinery. The ability to drive or to operate heavy machinery
may be impaired immediately after treatment.
• Only charge with the supplied cable from an USB-Socket with
5V dc and max 2500 mA.
• In the case of a defect please contact the manufacturer. Any
attempt to repair the device will void the warranty.
• Use with care as the cables may cause a tripping or strangu-
lation hazard.
• Don’t use in presence of ammable gases or in an oxygen en-
riched environment.
• Stop treatment and remove the electrodes if any irritation,
sensation, allergic reaction or burning occurs at the electrode
contact points and contact us.
• Take care that the units output connectors are always point-
ed towards the oor. This way the device is protected against
damage by water or liquids.
• Do not place electrodes on the chest; this can cause ventric-
ular brillation.
• The device is designed for indoor use only.
• The device should only be stored and transported under the
conditions dened in these instructions.
• Do not use the device in shielded locations, such as intensive
care units in hospitals, airplanes or wherever the use of mobile
phones is prohibited.
• The device has no life-sustaining functions.
• The device must be operated only with supplied accessories
and together with compatible smartphones.
• Not for use by those with a pacemaker or heart disease.
• Fixed, portable, mobile HF-communication devices and
RF-emitters, e.g. mobile phones, RFID, diathermy and securi-
ty systems (e.g. electromagnetic anti-theft systems, metal de-
tectors), may inuence medical electrical devices. You should
keep a safety distance of 30 cm (12 inchs) from HF-communica-
tion devices and RF-emitters of parts and cables of the Healy,
when the device is in use. Don’t stack it with such devices or
put it next to other electrical equipment when used.
• Avoid damages to the electrode cables and the charging cables
caused by sharp objects. If the cables are damaged do not use
them and contact the manufacturer for a replacement. Don’t
use cables that are damaged.
• To protect the device and its accessories against contamina-
tion and damage keep it closed and away from children, dust,