Use Instructions

16 | EN Healy GmbH | Schloss Kränzlin | Darritzer Str. 6 | 16818 Knzlin – Germany
7.2.3 Electromagnetic Compatibility
CISPR 11 -
emissions
EN55022
Group 2,
30 MHz – 1 GHz and 1 GHz – 6 GHz, classe B
The Healy is suitable for use in all establish-
ments, including the domestic area and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic
emissions
Class A acc. to IEC61000-3-2
Flicker Complies IEC61000-3-3
Immunity Complies IEC6061-1-2:2014 and EN61000-4-3
80 MHz – 2,7 GHz, 10 V/m
Immunity against wireless communication
equipment
28 V/m for MHz: 450, 810, 870, 930, 1720, 1845,
1970, 2450
27 V/m for MHz: 385
9 V/m for MHz: 710, 745, 780, 5240, 5500, 5785
ESD
protection
Complies IEC61000-4-2
8kV contact - 15 kV air
Power
frequency
magnetic
elds
Complies IEC61000-4-8
Test level 30 A/m for 50 Hz and 60 Hz
7.2.4 Radio Frequency Wireless Technology
Description Bluetooth Low Energy RF – Bluetooth V4
Compliance SIG TCRL Core - 2015-12-22; Core and Host Ver-
sion 2.0-4.2
(SIG = Bluetooth Special Interest Group)
ETSI EN301 489-1; ETSI EN301 489-17
Compati-
bility
Smartphones with Bluetooth Wireless techno-
logy and Bluetooth 4.0 or higher
Operating
frequency
2.4GHz
Congura-
tion
The Healy device starts the Bluetooth wireless
technology automatically when it is switched
on. Switch on the Bluetooth wireless techno-
logy on your smartphone and start the Hea-
ly App. During the rst scan, the app scans
and connects to the Healy device, see Section
3.1.5. The rst connection is called pairing.