2. Specifications
2.1. Technical Specification
Range
Current
Accuracy
50, 100, 150, 200µΩ 200A 0.8%
0.5, 1, 1.5, 2mΩ 100A 0.5%
5, 10, 15, 20mΩ 30A 0.2%
50, 100, 150, 200mΩ 10A 0.1%
0.5, 1, 1.5, 2Ω 3A 0.1%
The currents shown are the continuous rated currents for both AC and DC.
Higher currents can be used intermittently as supplied by pulse driven instruments. With an
ON to OFF time ratio of 1:10 or less, the allowed peak currents are 10x those specified
above, with an upper limit of 1000A and with a maximum applied time of 4 seconds. At least
1 minute should be allowed between applications for cooling.
It is important to ensure there is adequate low resistance connections on the 5070’s
current terminals.
2.2. General Specification
Dimensions: W540 x H210 x D410mm
Weight: 11kg