4.7. Opera ti on a s a Curre nt Source
As a current source the 1077 will
supply the set cur rent to the load
1. Check the setting of the output volt age control as in Sec t ion 3.1.
2. Turn on and select ‘Source’ mode.
3. Select most suitable range and set output on digit switches.
4. Connect ‘Load’ and turn output switc h O N.
Note: If V limit LED illuminates the 1077 has insufficient drive voltag e t o output the current
set. See ‘Output Voltage Control’ Section.
I