8
4. Calibration
4.1. Calibrating the 1017
The 1017 can be calibrated with a high accuracy D.M.M. with D.C. voltage, current and
resistance ranges with a specification of at least 4 times greater than that of the 1017’s
specifications.
Calibration is best carried out on fully charged batteries without the mains supply connected
to ensure that no mains supply interference takes any effect on the readings.
4.2. Zero Adjustment
Range Allowance Adjust Pot
1V 6µV 1V ZERO ON MODULE