Instruction Manual

4
1017 Technical Manual
4
1.2. Specifications
Outputs: 0 100V in 5 Ranges
0 100mA in 1 Range
0 10k in 1 Range
Voltage
Range: 0 9.99999mV in 10nV steps
0 99.9999mV in 100nV steps
0 999.999mV in 1uV steps
0 9.99999V in 10uV steps
0 99.9999V in 100uV steps
Accuracy: 10mV range 0.02% of setting +/- 0.005% of range
100mV range 0.01% of setting +/- 0.004% of range
1V range 0.005% of setting +/- 0.002% of range
10V range 0.005% of setting +/- 0.002% of range
100V range 0.01% of setting +/- 0.004% of range
Output Resistance: 100mV to 10V <200m
10mV and 100V <1
Output Current: 100mV to 10V 150mA
100V 10mA
10mV Limited by 1 Output resistance
All specification +/- 3uV.
Current
Range: 0-99.9999mA in 100nA steps
Accuracy: 100mA range 0.02% of setting and 0.004% of range
Drive Voltage: 10V Maximum
Resistance
Range: 0 9.99999k in 10m steps
Accuracy: 0.05% of full scale
Power Rating: 250mW per Resistor
End Resistance: < 200m
Deviation Control (Only on Voltage and Current)
Range: 0.001% to 0.999% in 0.001% steps
Accuracy: 0.5% Accuracy
Stability
Less than 10ppm/°C on all ranges.
Less than 5ppm/Day at constant temperature
Less than 15ppm/90 days
Less than 25ppm/Year
Output Noise
Output Noise (0.1Hz to 10Hz) - Typically 3ppm.
Power Supply
The 1017 can be used continuously from a 220V / 240V 50 / 60Hz (110V / 120V to order)
mains supply or from the internal rechargeable Ni-Cad battery pack. A front panel meter
indicates the state of charge at all times.