6000counts Operation Manual of Pocket Type Intelligent Scanning Digital Multimeter Index 1. Overview………………………………………………………………………………………………………………………..1 2. Safety precautions……………………………………………………………………………………………………………….1 3. Characteristics…………………………………………………………………………………………………………………...1 4. Operation panel instructions……………………………………………………………………………………………………3 5. Technical Specifications……………………………………………………………………………….…………...…….........4 6. Automatic switch machine………………………………………………………………………………….…………...……..8 7.
6000Counts Pocket Type Intelligent Scanning Digital Multimeter 1. Overview 6000counts is a pocket type 3 5 / 6-digit pocket type real effective value intelligent scanning digital instrument; the series is triple display: main display, secondary display, analog bar. Except for the display frequency of AC voltage measurement sub display, other function sub display is normal temperature. In the process of using, users can operate continuously. There is no need to turn the dial to select the function.
6000Counts Pocket Type Intelligent Scanning Digital Multimeter 1-9. Power: AAA 2* 1.5V battery; 1-10. Volume (size): 139×68×18mm (L×W×H); 1-11. Weight:about 175g(including 1.5V battery); 1-12. Attachment: One instruction manual, one certificate, one outer packaging box, one pair of test leads, two AAA1.5V batteries. 2. Technical characteristics 2-1.
6000Counts Pocket Type Intelligent Scanning Digital Multimeter 4. Operation panel instructions 1. Simulation bar; 2. LCD display; 3. Live wire simulation indication; 1 4. Normal temperature display status; 5. Quick selection of capacitor/flashlight key; 6. Data lock key; 7. Power on and off for more than 3 seconds; short trigger for AC voltage/ 2 3 DC voltage / diode / buzzer / capacitor switching, so as to cycle; 8. Electric field measurement (NCV) / neutral wire measurement (LIVE)/more than 3s℃/℉ 9.
6000Counts Pocket Type Intelligent Scanning Digital Multimeter 5.Technical Specifications 1.DC voltage / AC voltage automatic scanning test (DCV / ACV) Accuracy Range 6000counts Resolution DC/AC6V DC/AC60V 0.001V 0.01V ±(0.8%+5) DC/AC600V DC1000/AC750V 0.1V 1V ±(0.8%+10) Input impedance: 10MΩ; overload protection: 1000V DC or 750V AC peak.ture Validity Measurement: frequency response 50Hz-1kHz; The specific operations are as follows: 1-1.
POWER > 3S 6000Counts Pocket Type Intelligent Scanning Digital Multimeter 600kΩ 100Ω 6MΩ 60MΩ 1kΩ 10kΩ ±(2.5%+3) Input impedance: 10MΩ; overload protection: 1000V DC or 750V AC peak. The specific operations are as follows: 2-1. The boot display is in the automatic scan status "AUTO". 2-2. The black test lead is negative, the red test lead is positive, and the test lead reliably touches the test point. 2-3.
6000Counts Pocket Type Intelligent Scanning Digital Multimeter 4. Capacitance(C) Accuracy Range 6000counts Resolution 10nF 10pF 100nF 1uF 10uF 100pF 1nF ±(3.5%+20) 10nF 100uF 100nF 1mF 1uF 10mF 60mF 10uF 100uF ±(5%+3) Overload protection: 1000V DC, 750V AC peak. 4-1. The boot display is in the automatic scan status "AUTO". 4-2. The black test lead is negative, the red test lead is positive, and the test lead reliably touches the test point. 4-3.
6000Counts Pocket Type Intelligent Scanning Digital Multimeter 5. NCV measurement; The operation is as follows: 5-1. Press and hold the “NCV/LIVE” button; enter the EF measurement. 5-2. The front end of the meter has NCV test points. As long as the point is close to the AC voltage, the buzzer will emit different continuations according to the different strength of the signal, and the LCD will also display according to the strength of the signal. Different number of segments. 5-3.
6000Counts Pocket Type Intelligent Scanning Digital Multimeter 3) When measuring high voltage circuits, special care should be taken to avoid electric shock; 4) It is forbidden to input voltage values exceeding 250V DC or AC peak to avoid damage to the meter. 6.