Passive Isolating Optical Table Supports
5) Repeat items (2) to (4) until all four isolators have a gap of 63 mm (2.48”) between the top of the cylinder and the
underside of the table – see Fig. 3.4.
Fig. 3.4 Correct inflation of passive isolators
6) Check that the table top is level. If the floor is uneven, the gap between the table and each isolator may not be the same.
If this is the case, identify the isolator with the smallest gap. Set this gap to 63 mm (2.48”) and then adjust the remaining
isolators such that the table is level. Use a levelling device as necessary.
63 mm (2.48")
63 mm (2.48")
under inflated
over inflated
correctly inflated
standard load
heavy load