User Guide

6 HA0206T Rev F Sep 2022
Chapter 3
3.2.3 Mounting the Table on the Isolators
1) Raise the table on a fork-lift truck, ensuring that the fork distance is set such that the forks can pass between the isolators
when the table is lowered into position – See Fig. 3.2.
Fig. 3.2 Distance between forks
2) Manoeuvre the fork-lift truck until the table is over the isolators. Carefully lower the table until it is around 20mm above the
isolators.
3) Lower the table onto the isolators.
4) When the table is supported on the isolators, remove the fork-lift truck.
3.3 Set-up
1) Remove the valve caps from each isolator.
2) Connect a footpump to the Schraeder valve of one of the isolators and inflate to about 40 p.s.i. - see Fig. 3.3
3) Repeat item 2) for the isolator diagonally opposite.
4) Repeat items (2) and (3) for the remaining two isolators.
Fig. 3.3 Inflating passive isolator
forks
distance between isolators