Owner's manual

30
31
TR-0206 Stainless Protection Sensor
Cable Length : 0.6m
Thermal-Constant Time :
In the air : Approx. 75 Sec.
TR-0306 Stainless Protection Sensor
Cable Length : 0.6m
Thermal-Constant Time:
In agitated water : Approx 18 Sec.
Only stainless section is water resistant
TR-0406 Stainless Protection Sensor
Cable Length : 0.6m
Thermal-Constant Time :
In agitated water : Approx 20 Sec.
Only stainless section is water resistant
TR-0506 Stainless Protection Sensor
Cable Length : 0.6m
Thermal-Constant Time :
In agitated water : Approx 20 Sec.
Only stainless section is water resistant
TR-0706 Stainless Protection Sensor
Cable Length : 0.6m
Thermal-Constant Time :
In agitated water : Approx 18 Sec.
Only stainless section is water resistant
Materials:
Thermistor
TPE resin-shielded sensor
TPE resin-shielded wire
M3Screw Hole
Compacion Tube
Stainless pipe
(
SUS304
)
Stainless pipe
(
SUS316
)
Possible Measurement Range :
-
40 to 110
℃  
Sensor Temperature Durability:
-
50 to 115
Measurement Accuracy : Average ±0.3
(-
20 to 80
)
Average ±0.5
(-
40 to
-
20
/
80 to
110
)
Sensor Extension CableTemp Sensor Only
unit : millimeters
TR-1C30
Cable length : 3.0m
Materials:
Vinyl Coated Electrical Wire
Note:
- Only one cable per sensor. When using the extension cable there will be
a +0.3 at normal temperature and at -50 a gap of +0.5 may occur.
- If a sensor extension cable is being used with the data logger connected
by USB to your computer, electromagnetic waves may cause large errors
in measurements.
- Do not use the Extension Cable with TR-72U.
Temp / Humidity SensorsFor TR-72U
TR-3100 Temp / Humidity Sensor
TR-3110 Temp / Humidity Sensor
Cable length : 1m
Materials:
Temperture/humidity sensor
Polypropylene resin
Vinyl Coated Electrical Wire
Possible temperature measurement range :
0 to 55
Possible humidity measurement range :
10 to 95% RH
Sensor temperatur resistance :
-
10 to 55
humidity measurement accuracy :
±
5
RH
At 25
50
RH
Service life :
1 year (under normal operational conditions)
Operational conditions : Without dew condensation, water leakage or effect from
corrosive gas or organic solvents.
Note: An Extension Cable cannot be used with Temperature/Humidity
Sensors.