Configuration Guide User Manual
A Note about Microsoft Exchange Server: The TR-700W will work with MS Exchange Servers. In order to do so, however, it will
be necessary to have the IT Specialist who maintains your server set up an unencrypted Client to Server Connector for the TR-700W
on port 587
If you have elected to enable Warnings you will want to
set the limits which trigger them. You can set upper
limit or a lower limit, or both for either channel or both
channels.
Note that due to the internal workings of the logger the
temperature limits for Warnings must be stated in
Degrees Celcius.
You are also asked to enter a Warning Judgment Time.
This is not the warning sampling Interval. This time is
the duration that the set limit must be exceeded before a
warning notification will be sent. The unit will send a
Warning Notification to all designated recipients when
this condition is met. It will only send the warning once,
but will send an OK Notification when the readings
return to within normal limits.
The last entry on this page enables the unit to send a Warning Notification when the unit detects that a sensor has been removed or has
become defective. You can enable this for either channel or for both.