OPERATION MANUAL MULTI-FUNCTION DATALOGGER WITH PRINTER MULTI-FUNCTION DATALOGGER WITHOUT PRINTER WITH PRINTER 9811 9812 9861 9871 9881 9882 9851 9680 9831 98315 9832 9833 9835 WITHOUT PRINTER 9611 9612 9661 9671 9681 9682 9651 9660 9631 96315 9632 9633 9635
INDEX Introduction ............................... Material Supplied ......................... Features ................................... Comparison table ......................... Meter keypad - with printer .............. Meter keypad - without printer .......... Accessory photo .......................... Top cover of meter ........................ Meter setting ............................... 1 1 1 2 5 6 7 8 10 Single measurement ....................... 11 Multiple measurement .....................
INTRODUCTION Thank you for purchasing this multiple function printer! This unit has been developed to meet your max.satisfaction by its userfriendly design. Please review the entire manual for a complete overview of how to operate this meter. MATERIAL SUPPLIED Check for damaged or missing parts in your meter before starting. The complete meter set should contain 1. Meter 2. 4pcs AA batteries 3. Printing Thermo-paper ( For printer models only) 4. Operation manual 5. Carry Case 6.
COMPARISON TABLE Measured parameters of each meter: MODEL FUNCTION PARAMETERS Temperature 9811/ 9611 IR Thermometer 9812/ 9612 6 IR + external probe Meter 9861/ 9661 pH meter pH value, Voltage Temperature 9871/ 9871 Anemometer Velocity, Humidity Temperature Wet Bulb, Air Volume 9881/ 9681 Dual K Type Thermometer Temperature 9882/ 9682 Dual K.J.T.R.S E Type Temperature Diff.
Please refer to below for the difference of each measurement mode.
Comparison table of multiple points and datalogging measurements MULTIPLE POINT DATALOGGING MEASUREMENT RECORD At most 12000 points 99 points (Note 4) Editable or FILE Default with date NAME and time (Note 1) Press "ON/OFF" key to measure and store MEASURING / by pressing specified SAMPLING keys Default (Note 2) Automatically measure and store according to pre-set parameters (Note 3) Note 1: Each record is designed to show with current date and time if user doesn't edit any.
METER KEYPAD-WITH PRINTER Top Cover Based on different functions, the top cover connector and color are different. Welcome MEAS F1 MEM F2 LOG LCD SET F3 Corresponding Function keys F4 1 SYM. 2 ABC DEF 3 4 GHI 5 JKL MNO 7 PQRS 8 TUV WXYZ 9 6 0 ON/OFF LINE FEED Keypad 1SYM stands for"1,*,.
METER KEYPAD-WITHOUT PRINTER Top Cover Based on diff. functions, the top cover connector and color are different. METER KEYPAD Big dot matrix LCD Welcome Corresponding Function keys Press F1~ F4 key to select the needed function MEAS MEM 1 SYM. SET F3 F2 F1 LOG 2 ABC F4 DEF 3 4 GHI 5 JKL MNO 7 PQRS 8 TUV WXYZ 9 6 0 To view previous /next record ON/OFF LINE FEED Keypad 1SYM stands for"1,*,.
pH connector Temp. connector ACCESSORY 9861 9661 9871 9671 When first time using or the probe hasn't been used for a long time, please calibrate the probe first by following the procedure in page 20. Please always keep the pH glass bulb wet by using the cap to protect and store the electrode. The temperature and humidity sensors are built in the fan, so remember to rotate the cover before using to get accurate reading. Always rinse the pH eletrode in de-ionized water or rinse solution (tap water.....
TOP COVER -CONNECTOR The following illustrations are the connectors for different printers: 9871/9671 Anemometer 9851/9651 Hygro-thermometer 9811/9812/9611/9612 Infrared Thermometer 9832/9835/98315/9833/9831 9632/9635/96315/9633/9631 9881/9681 K Thermometer 9882/9682 KJTRSE thermometer Manometer 9861/9661 pH meter 9680/9660 IrDA printer REMARKS: Welcome any OEM for different kind of parameters . 9821 Sound Level Meter connector's illustration is provided when require .
9680/9660 IrDA RECEIVING METER An IrDA like module is built in the meter 9680/9660 at the top , better target the IrDA within 30 degree angle toward any of AZ IrDA meters for receiving data . Make sure you have disabled Auto power off from both AZ IrDA meter (transmitter )and IrDA receiver when you need to have long-term measurement.
METER SETTING Before measuring, please check the meter setting first to confirm the basic settings of the meter are what you need. To enter each command, just press the corresponding F1 to F4 key. Welcome MEAS F1 MEM F2 LOG F3 SET F4 SET F4 :Meter Setting Pressing or key to shift the cursor. Pressing EDIT to enter modification mode. Pressing PRN to enter printing mode. Pressing NEXT/BACK to enter next or previous page. Pressing EXIT to return to main menu Descriptions : LCD Cont.
SINGLE MEASUREMENT There are three measurement modes : 1. Single measurement(MEAS) 2. Multiple measurement(MEM) and 3. Auto logging (LOG). Operating the single measurement as a general meter. MEAS F1 :Single Measurement Welcome MEAS F1 MEM LOG F2 F3 SET F4 Definition Press ON/OFF key to measure the temperature for 9811 and 9611( No printer function ). Press F1 (MEAS) to proceed single measurement for other meters . On 9812/9612, the external probe temp.
MULTIPLE MEASUREMENT This is to manually record what you measured with real time , and file name (a note or identification for the record ) is editable to memorize or recognize . There are total 99 memorize capacities in this measurement. MEM F2 :Multiple measurement Welcome MEAS F1 MEM F2 LOG F3 SET F4 Definition Press ON/OFF ( IR models 9811/9611/9812/9612) or MEAS (other models) to measure the parameters. Select F1(ABORT) or F4(SAVE) to exit the measurement or memory the record .
AUTOMATIC LOGGING This is to automatically record what you set up before measurement with real time, first to set records start/stop date and time, sample rate , memory points. The screen content shows Expect memory points and Remain memory points while in setting. LOG F3 :Datalogging Welcome MEAS F1 MEM LOG F2 F3 SET F4 Definition Press SET to set needed setting first. The Date mode is based on the mode you set under SET (see page 10). The sampling rate is from 1 to 7200 seconds.
9811/9611 Infrared thermometer Setting / Single measurement Welcome MEAS MEM LOG Emi. Rate: 0.95 LCD Cont.(1-5): 3 Prn Cont. (1-9): 5 Unit: C Laser: Enable Auto Off: 20mins Enable EXIT EDIT PRN T:30.5 C Press ON to Measure EXIT SET PRN NEXT Set Clock: MM-DD-YY HH:MM:SS 06 - 30 - 05 18 : 29 :2 0 Set ID: Disable ID: User Name Print? Press ON to Measure YES NO EXIT EDIT PRN BACK Cont 5 Press ON to Measure STOP HIGH LOW Remarks: 1.No PRN function is existing for model 9611 IR meter. 2.Emi.
9812/9612 IR meter w/ probe Setting / Single measurement Welcome MEAS MEM LOG Emi. Rate: 0.95 LCD Cont.(1-5): 3 Prn Cont. (1-9): 5 Unit: C Laser: Enable Auto Off: 20mins Enable EXIT EDIT PRN Ext:30.5 C Ir: Press ON to Ir Meas CALI EXIT SET PRN NEXT Press on/off key Set Clock: MM-DD-YY HH:MM:SS 06 - 30 - 05 18 : 29 :2 0 Set ID: Disable ID: User Name Ext:30.5 C Ir: 29.8C Press ON to Ir Meas EXIT CALI PRN EXIT EDIT PRN BACK Remarks: 1.No PRN function is existing for model 9612 IR meter. 2.
9811/9611 Infrared thermometer Multiple measurement Welcome MEAS 01:06-30 02:28:29 T:-0.3C 02:06-30 02:28:45 T:26.7C 03:06-30 02:28:55 T:111C EDIT LOG SET 01:06-30 02:28:29 T:-0.3C 02: Press ON to Measure EXIT MEM CLR PRN 01:06-30 02:28:29 T:-0.3C 02:USER name 1111 T:26.
9812/9612 Infrared thermometer Multiple measurement Welcome MEAS 01:06-30 02:28:29 Ext:-0.3C Ir: 15.0C 02:06-30 02:28:45 Ext: Ir: Press ON to Measure EXIT EDIT CLR PRN 01:06-30 02:28:29 Ext:-0.3C Ir: 15.0C 02: USER name 1111 Ext: Ir: MEM LOG SET 01:06-30 02:28:29 Ext:-0.3C Ir: 15.0C 02:06-30 02:28:45 Ext: Ir: Press ON to Measure EXIT EDIT From Point: From Point: CLR PRN 1 99 Press ON to Measure EXIT EDIT CLR PRN Short press ESC EDIT START Long press 01:06-30 02:28:29 Ext:-0.
9811/9611/9812/9612 IR Thermometer Datalogging measurement Welcome MEAS Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 12000Point(s) Remain: 11997Point(s) EXIT EDIT VIEW Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 12000Point(s) Remain: 11997Point(s) START CLR PRN MEM LOG SET 00001: 06-30 02:28:55 T:-0.3C 00002: 06-30 02:29:55 T:-0.5C 00003: 06-30 02:30:55 T:-0.
9812/9612 External Temp. Probe Calibration procedure Welcome MEAS MEM LOG SET Ext:30.5 C Ir: Press ON to Ir Meas CALI EXIT PRN Long press Probe Cali Data Step 1 :0 at 5C Step 2: -5 at 40C Step 3: -6 at 85C Note: Unit is 0.1C LXW SAVE EDIT EXIT Input Cali Data Step 2 :+ 00 ABORT LXW +/- SAVE ENTER Repeat until three points value are saved Input Cali Data Step 1 :+ 00 ABORT LXW +/- SAVE ENTER Remarks: 1.
9871/9671 Anemometer +Psychrometer Setting/Single measurement Welcome MEAS T: 29.7C V: WBT: LOG RH: 73.7% VOL: 0.1CMM INPUT EXIT PRN LXW D EXIT AREA 000.0 cm ESC NEXT EDIT PRN NEXT BACK ENTER T: 29.7C RH: 73.7% Print? WBT: 26.1C Remarks: PRN Select Item MPS V C T % RH C WBT CMM VOL EDIT PRN EXIT ~ Range is 0 300cm L: EDIT Set Clock: MM-DD-YY HH:MM:SS 06 - 30 - 05 18 : 29 :2 0 Set ID: Disable ID: User Name Please select type ESC SET LCD Cont.(1-5): 3 Prn Cont.
9871/9671 Anemometer +Psychrometer Multiple meas/Datalogging Welcome MEAS 01: V: T: RH: WBT: VOL: MEAS EXIT 01: T: 29.7C V: WBT: 26.1C EDIT MEM NEXT RH: 73.7% 0.7MPS VOL: 0.1CMM ABORT SAVE LOG SET 0001: V: T: RH: WBT: VOL: START SET EXIT NEXT Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 2400Point(s) Remain: 2400Point(s) EXIT EDIT VIEW NEXT Press START to begin 01: 06-30 23:56:49 0.7MPS V: 29.7C T: 73.7% RH: WBT: 26.1C 0.
9881/9882/9681/9682 Thermometer Setting/Single measurement Welcome MEAS T2: 29.7C T1: 39.7C WBT: EXIT TYPE K T1-T2:: 10.0C EXIT NO T2: 29.7C T1: 39.7C WBT: TYPE K EDIT PRN NEXT Set Clock: MM-DD-YY HH:MM:SS 06 - 30 - 05 18 : 29 :2 0 Set ID: Disable ID: User Name Print? YES SET LCD Cont.(1-5): 3 Prn Cont. (1-9): 5 Auto Off: 20mins Enable 9881 T1-T2:: 10.0C 29.7C EXIT LOG PRN T2: WBT: MEM EXIT T1-T2: 10.
9881/9882/9681/9682 Thermometer Multiple measurement /Datalogging Welcome MEAS 01: T1: T2: T1-T2: 02: T1: T2: MEAS EXIT 01: T2: MEM LOG SET 0001: 0002: EDIT 86.1F T1: 56.1F ABORT 01: 06-30 23:56:49 56.1F T1: 86.1F T2: T1-T2: 30F 02: T1: T2: MEAS EDIT EXIT 01: 06-30 23:56:49 56.1F T1: 86.
9861/9661 pH/mV Setting / Single measurement Welcome MEAS Volt: -34.9mV PH: 7.60 MEM LOG Temp: 28C LCD Cont.(1-5): 3 Prn Cont. (1-9): 5 Unit: Metric Date: YY-MM-DD Auto Off: 20mins Enable EXIT EDIT PRN ATC CAL EXIT SET PRN NEXT Long press Set Clock: MM-DD-YY HH:MM:SS 06 - 30 - 05 18 : 29 :2 0 Set ID: Disable ID: User Name "CALIBRATION" -3.6mV 7.00 Auto Judging...
9861/9661 pH/mV Probe Calibration procedure-1 Welcome MEAS Volt: -34.9mV PH: 7.60 MEM Temp: 28C ATC CAL EXIT PRN LOG SET Note This is to calibrate under ATC( Temperature probe is plugged side by pH probe ) To calibrate under NonATC, please see procedure -3) Long press CALIBRATION -3.6mV 7.00 Auto Judging... Last: 76 days ago CALIBRATION Volt: PH: EXIT LXW RST SAVE ENTER PH: 7.00 Adj pH Value ATC EXIT CALIBRATION LXW RST SAVE ENTER SAVE ENTER CALIBRATION PH: 7.
9861/9661 pH/mV Probe Calibration procedure-2 Continuing from last page Ex. PTS's value is 75% CALIBRATION PTS: 75.0% TEMP: 27.0C CALIBRATION PTS: 100.0% TEMP: 27.0C ATC ATC EXIT LXW RST SAVE ENTER EXIT Back to Auto judging next point cal. Quit with save? Quit without save? ( Quit with save? ) YES NO NO Press EXIT ,"Quit without save?" is appeared; Press SAVE,"Quit with save ?" is appeared.
9861/9661 pH/mV Probe Calibration procedure-3 Welcome MEAS Volt: Temp: -34.9mV PH: 7.60 28C NonATC EXIT TEMP CAL PRN RANGE:-5.0~80.0C o Temp: 18.2 C EXIT LXW CLR +/- MEM LOG SET Note 1: This is to calibrate under NonATC ( No Temperature probe is plugged side by pH probe ) Note 2: If you always see "--.--" pH in autojuddge, please refer to trouble shooting. Note 3: Display turns to normal mode if non-operation during Temp. is flashing for editing. Volt: Temp: 28C -34.9mV PH: 7.
9861/9661 pH/mV Multiple measurement Welcome MEAS 01: PH: Volt: Temp: 02: PH: Volt: MEAS EXIT 01: Volt: -35.6mV PH: 7.49 EDIT MEM 01: 23:56:49 01: 06-30 PH: 7.49 Volt: -35.6mV Temp: 86.1F 02: PH: Volt: MEAS EDIT EXIT SAVE SET 01: PH: Volt: Temp: 02: PH: Volt: CLR EXIT NEXT Temp: 86.1F ABORT LOG PRN BACK Note: After saving #01 memory, press for next measurement #02 with a cursor will be highlighted. 01: 23:56:49 01: 06-30 PH: 7.49 Volt: -35.6mV Temp: 86.
9861/9661 pH/mV Datalogging Welcome MEAS 0023: 06-30 02:28:55 PH: 4.88 Volt: 127.0mV Temp:28.8C 0024: PH: Volt: EXIT START SET NEXT Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 4000Point(s) Remain: 3977Point(s) EXIT EDIT VIEW NEXT Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 4000Point(s) Remain: 3977Point(s) START CLR PRN BACK LOG SET 0023: 06-30 02:28:55 PH: 4.88 Volt: 127.0mV Temp:28.
9680/9660 IrDA Setting and Single Data Receiving Welcome MEAS MEM LOG NEXT Welcome Welcome MEM MEAS LOG NEXT PRN SET BACK (Take 8726 as an example) RH: Air: 24.9 C 60.1% T1: 30.5 C DP: 16.5 C 19.5 C PRN Air: RH: 24.9 C Print? Enable WBT: EXIT DP: LCD Cont.(1-5): 3 Prn Cont. (1-9): 5 Auto Off: 20mins 60.1% EXIT EDIT PRN NEXT Set Clock: MM-DD-YY HH:MM:SS 06 - 30 - 05 18 : 29 :2 0 Set ID: Disable ID: User Name WBT: 16.5 C YES 19.5 C NO EXIT EDIT PRN BACK REMARKS: 1.
9680/9660 IrDA Multiple and Datalogging data Receiving Welcome MEAS MEM MEAS EDIT EXIT NEXT CLR PRN START SET Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 4000Point(s) Remain: 3977Point(s) EXIT EDIT VIEW 01: 08-05 15:13:50 Air: 24.9C RH: 60.1% DP: 16.5C WBT: 19.5C T1: 30.5C EXIT NEXT 01: 08-05 15:13:50 Air: 24.9C RH: 60.1% DP: 16.5C WBT: 19.5C T1: 30.5C 01: 08-05 15:13:50 Air: 24.9C RH: 60.1% DP: 16.5C WBT: 19.5C T1: 30.
9680 Receiving & Printing up to 99 points at one time IrDA meter example: 8726 Psychrometer Welcome MEAS MEM LOG NEXT Welcome PRN SET BACK Expect: 99 Received: 0 Remain: 99 EXIT After receiving data from IrDA meter Expect: 99 Received: 31 Remain: 68 EXIT Automatically start to print REMARKS: 1. To set the IrDA meter at printing mode otherwise above procedure could not be processed. Please refer 5 Cont to the manual of each IrDA meter for the detail printing STOP HIGH LOW mode setting method. 2.
9680/9660 IrDA Multiple data Receiving with IrDA meters There is no problem for single data transmitted from IrDA meter to IrDA receiver even you change the meter frequently. However it might happen " Inconsistent " if you try to change the IrDA meters from one model to the other model during the ] Multiple and Datalogging IrDA receiving . No IrDA meter is targeting to the IrDA data receiver.
9680/9660 IrDA Datalogging Receiving with IrDA meters No IrDA meter is targeting to the 9680/9660 data receiver. (Press LOG w/o IrDA meter) 01: 08-05 15:13:50 Air: -------RH: -------DP: -------WBT: -------T1: -------EXIT START Press START SET NEXT Logging ...... STOP MEAS VIEW Though you might not target 9680 to an IrDA meter yet, when you press START(F2), Logging will still be activated. The IrDA receiver is proceeding receiving per presetting sampling time.
983X /963X Manometers Setting / Single measurement Welcome MEAS MEM LOG Temp: LCD Cont.(1-5): 3 Prn Cont. (1-9): 5 Auto Off: 20mins Enable P: 0.1psi EXIT REC Cali PRN EXIT Max: Min: Avg: Curr. LXW HI LOW Pressure EDIT PRN NEXT Set Clock: MM-DD-YY HH:MM:SS 06 - 30 - 05 18 : 29 :2 0 Set ID: Disable ID: User Name CALIBRATION ESC SET ZERO EXIT EDIT PRN NEXT Temp: 1.2psi -1.2psi 0.4psi 0.1psi Select Item P psi EXIT EXIT Remarks: EDIT PRN BACK 1.
983X / 963X Manometer Calibration Procedure Welcome MEAS MEM LOG SET Temp: Range: xx ~ xx psi P: 0.1psi EXIT REC Cali Cali: * * * PSI PRN ESC LXW (remak3) SAVE ENTER Long press Cali: * * * PSI CALIBRATION (remak3) Wait ESC LXW HI LOW ZERO After 5 seconds Range: xx ~ xx psi (remark2) Cali: * * * PSI ESC LXW Cali: * * * PSI (remak4) (remak1) OK ( or NG) SAVE ENTER EXIT Press up/down or number keys to adjust 1.There are up to three calibration points.
While in calibration HI mode, the default HI value will be displayed on LCD as 1.600 or 4.000... based on different models. While in the calibration LOW mode, the default LOW value will be displayed as well. 2. If your standard calibrator is not the same as meter default value, press up/down or number keys to adjust. The adjustable range is displayed on LCD as a reminder. Please refer to table in page 36 for the details. The ZERO calibration could not be adjusted and the value always should be 0.000 3.
983X/963X Manometer Multiple measurement Welcome MEAS MEM LOG 01: SET 01: P: 02: 03: P: 02: P: 03: P: EXIT MEAS EDIT NEXT 01: P: 0.100psi ABORT 01: 01: 06-30 23:56:49 P:0.100psi 02: P: 03: P: MEAS EDIT EXIT SAVE P: P: EXIT CLR PRN BACK Note: After saving #01 memory, press for next measurement #02 with a cursor will be highlighted. 01: 01: 06-30 23:56:49 P:0.100psi 02: P: 03: P: CLR PRN EXIT NEXT BACK Remarks: 1.No PRN function is existing for model 963X manomenters. 2.
983X/963X Manometer Datalogging Welcome MEAS 0023: 06-30 02:28:55 P: 0.1 psi 0024: P: 0025: P: EXIT START SET Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 12000Point(s) Remain: 11977Point(s) EXIT EDIT VIEW Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 12000Point(s) Remain: 11977Point(s) START CLR PRN NEXT MEM LOG SET 0023: 06-30 02:28:55 P: 0.1 psi 0024: P: 0025: P: P-PG N-PG PRN BACK 0023: 06-30 02:28:55 P: 0.
9851 /9651 Hygo-Thermometer Setting / Single measurement Welcome MEAS LOG DP Temp: RH 60.8% WBT MEM SET 17.4C LCD Cont.(1-5): 3 Prn Cont. (1-9): 5 Auto Off: 20mins Enable PRN EXIT Temp:25.5C 20.0C EXIT REC Cali LXW INPUT EXIT MANUAL Temp REC Max : 20.5C Min: 20.0C Avg: 20.2C Curr.: 20.3C EXIT EDIT Temp: PRN NEXT NEXT Select Item Temp RH DP WBT NEXT EXIT RH REC Max : 60.5% Min: 60.0% Avg: 60.2% Curr.: 60.
9851/ 9651 Hygro-Thermometer Calibration Procedure Note: "Input cali.data" method is for RH probe replacement. "Manual calibration" method is for RH probe replacement & old probe re-calibration. Welcome MEM MEAS LOG SET (Need to plug probe into salt bottles) 60.8% WBT RH Manual Calibration DP Temp: RH 17.4C Temp:25.5C RH 60.8% Temp. 23.5 C 20.0C EXIT REC Cali EXIT PRN Long press RH Cali Mode Select Input probe cali data Press F2 Calibrate manually Press F3 EXIT LXW INPUT LXW RH 32.
9851/9651 Hygro-Thermometer Multiple measurement Welcome MEAS MEM 01: Temp: RH: DP: WBT: 02: EXIT MEAS EDIT NEXT 01: DP Temp: RH 60.8% WBT 17.4C Temp:25.5C 20.0C ABORT 01: SAVE MEAS EDIT SET 01: Temp: RH: DP: WBT: 02: EXIT CLR PRN BACK Note: After saving #01 memory, press for next measurement #02 with a cursor will be highlighted. 01: 06-30 23:56:49 Temp: 25.5C RH: 60.8% DP: 17.4C WBT: 20.0C 02: EXIT LOG NEXT ROOM1 Temp: 25.5C RH: 60.8% DP: 17.4C WBT: 20.
9851/9651 Hygro-Thermometer Datalogging Welcome MEAS START LOG SET 0001: Temp: RH: DP: WBT: 0002: 0001: Temp: RH: DP: WBT: 0002: EXIT MEM SET Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 3000Point(s) Remain: 3000Point(s) EXIT EDIT VIEW Begin: 30-06-05 Start: 02:28:55 End: 02-07-05 Suspend: 03:28:55 Rate: 60Sec(s) Expect: 3000Point(s) Remain: 3000Point(s) START CLR PRN NEXT P-PG N-PG PRN BACK NEXT 0001: 06-30 02:28:55 BACK Temp: 25.5C RH: 60.
SOFTWARE The enclosed software is a quick tool for you to download the memorized data to PC for further analysis or upload the pre-edited file names to meters, this may save your setting time. Material & O/S Software CD and RS232 cable with D-sub plug are needed. Operating system need to be WIN98/2000/NT/XP or above. Download Download Download Upload New Save Load Print Memory Logger file names All FILE New: To create a new file. Load: To open a saved file. Save: To save current file.
Print Load MODE There are two data modes. One is Memory, the other one is Logger. To select the mode from main menu or quick selection window. In logger mode, you may select from "Goto" to choose logged data range. There are 100 points in one page.
PORT Select the correct Com. port to ensure the meter is communicated with PC. I n connecting mode, "PC Mode" will appear on the LCD and "COM #" & Edit will show on the left bottom corner of screen. For most PCs, the Com. port number is COM1. There are 8 COM ports for selection in this software. Port COMMAND There are four commands in this software. -Download measured logging data from meter to PC. -Download measured memory data from meter to PC. -Download measured logging & memory data from meter to PC.
PRINT Select the data you want to print : There are 3 selection : 1. Print only Logger Data 2. Print only Memory Data 3. Print all ( Both Logger Data and Memory Data in sequel ) Example: (If you choose Print all from 9811 IR meter ) Page1 << DataLogging Printer Report Date:07-12-2005 TIME:17:26:30 >> Memory Data: STEP DESCRIPTION ======= ============== 1 07-12 10:46:24 2 07-12 10:46:30 3 07-12 10:46:33 4 07-12 10:46:38 5 07-12 10:46:48 ............................. Temp ====== 26.2 C 26.5 C 25.8 C 25.
98XX SERIES SPECIFICATION SHEET 50 48 10~90%RH 10~90%RH
96XX SERIES SPECIFICATION SHEET 49
MANOMETER SPECIFICATION SHEET TYPE 9832/9632 9835/9635 Unit psi Resolution psi 1 0.001 Inch of H2O 27.680517 0.01 bar 0.068948 0.001 mbar 68.948253 0.1 mm of Hg 51.712016 0.1 Inch of Hg 2.035907 0.01 oz/inch2 16.000844 0.02 Kg/cm2 0.070309 0.001 kPa 6.894859 0.01 Ft of H2O 2.306719 0.01 cm of H2O 70.309000 0.1 Resolution 98315/96315 9833/9633 9831/9631 Resolution Resolution Resolution 0.003 0.1 0.001 0.2 0.2 0.01 0.05 0.001 0.02 0.01 0.2 0.01 0.3 0.001 1 0.5 0.02 0.2 0.001 0.1 0.02 1 50 0.02 0.5 0.
TROUBLE SHOOTING For all datalogging printers: 1. Power on but no display a)Make sure the time of pressing "ON/OFF" key is more than 200 mini seconds(ms). b)Check the battery are in place and make sure they are at good contact and correct polarity. c)Replace with new batteries and try again . d)Move the batteries for ten seconds ,then replace back again. 2. Display disappear a)Check whether the low battery indicator is displayed on or before display disappears. If yes , replace with new batteries .
2) 3) 4) 5) b)PTS of some calibration points are out of 85%-105%. c)Replace with new batteries and solution, then try again. d)If it fails again, return the meter to the dealer for repairing. E2. Problem: The value is underflow. E3. Problem: The value is overflow. E4. Problem: The original data that relate to the value is error. E31. Problem: A/D measurement error. For 9680/9660 IrDA receiver and printer 1) E5 . Problem : The value iis exceeding 999999 or less than -99999.