User guide
b. Press SET then use the UP or DOWN keys until the correct numerical
setting is displayed. Press SET to accept the new value.
c. Repeat step b for ALPHA and DELTA.
10.2.3 Accuracy & Repeatability
1. Check the accuracy of the dry-well at various points over the calibrated
range.
2. If dry-well does not pass specification at all set-points, repeat the Calibra
-
tion Procedure.
45
10 Calibration Procedure
Calibration Procedure