2 Introduction
The Hart Scientific 9141 High-Temp Field Calibrator may be used as a portable
instrument or bench top temperature calibrator for calibrating thermocouple
and RTD temperature probes. The 9141 is small enough to use in the field, and
accurate enough to use in the lab. Calibrations may be done over a range of
50°C to 650°C (122°F to 1202°F). Temperature display of the 9141 is 0.1
degrees.
The dry-well calibrator features:
Rapid heating and cooling
Interchangeable multiple hole probe sleeves
Convenient handle
RS-232 interface
Switchable AC Input (115 VAC or 230 VAC)
Built in programmable features include:
Temperature scan rate control
Temperature switch hold
Eight Setpoint memory
Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s hybrid analog/digital con-
troller. The controller uses a precision platinum RTD as a sensor and controls
the well temperature with a solid state relay (triac) driven heater.
The LED front panel continuously shows the current well temperature. The
temperature may be easily set with the control buttons to any desired tempera
-
ture within the specified range. The calibrator’s multiple fault protection de
-
vices insure user and instrument safety and protection.
The 9141 dry-well calibrator was designed for portability, low cost, and ease of
operation. Through proper use, the instrument will provide continued accurate
calibration of temperature sensors and devices. The user should be familiar
with the safety guidelines and operating procedures of the calibrator as de
-
scribed in the instruction manual.