Specification Sheet
maximum
convenience for
maximum safety
13
SPECIFICATIONS TABLE/ORDER NUMBERS LARGE CAPACITY GENERAL PROTOCOL OVENS
Order number 51029327 51029339
Model OGS400 OGS750
Convection technology gravity convection gravity convection
Temperature range °C +50 °C
1
- 250 °C +50 °C
1
- 250 °C
Spatial temperature deviation at 150 °C ±3.0 °C ±3.7 °C
Temperature deviation over time at 150 °C ±0.5 °C ±0.5 °C
Footprint m
2
/ sqft 0.56 / 6.0 0.91 / 9.8
Chamber volume L / cuft 419 / 14.8 774 / 27.3
Dimensions chamber mm / in (W x H x D) 544 x 1307 x 590 / 21.4 x 51.5 x 23.2 1004 x 1307 x 590 / 39.5 x 51.5 x 23.2
exterior
2
mm / in (W x H x D) 778 x 1545 x 770 / 30.6 x 60.8 x 30.3 1261 x 1545 x 770 / 49.6 x 60.8 x 30.3
Number of shelves supplied / max 2 / 39 2 / 39
Max. shelf load kg / lb 40 / 88 40 / 88
Rated voltage / frequency V / Hz 208-240 / 60 208-240 / 60
Rated power / max. current W / A 2640 / 11 3240 / 13.5
Energy consumption at 150° C W 520 800
Weight kg / lb 136 / 300 182 / 401
maximum
convenience for
maximum safety
•
Large, easy-to-view, vacuum fluorescent display with simple-to-use touch button operation controlled
by an onboard microprocessor
•
Doors can be opened over 180 ° for ease-of-access
•
Easy-to-remove shelves for cleaning
•
Automatic overtemperature alarm system to protect samples – no need for timely manual setting
•
Timer function to program start or switch off times for additional energy savings
•
Lockable casters for easy mobility and stability
Easy-to-view, vacuum fluorescent display
with simple-to-use touch button operation
Factory installed casters allow for increased
mobility and make setting up your lab easy
Heratherm large capacity ovens offer up to 23%
energy savings compared to conventional ovens*
* based on internal testing 04/2012
1
Temperatures as low as ambient +10 °C can be selected – requires open damper and no additional heat in unit
2
Depth of handle / display not included in depth (2.6 in); casters not included in height (4.3 in) – required distance to rear wall: 4.7 in
NOTE: All figures in this table are typical values for series devices, based on factory
following norm DIN12880 (measured with 240V, values will differ with 208V). Please
contact your sales representative for certification information for IQ/OQ documents.