Specification Sheet
.1)
Optional stainless
steel exterior
• Robust and corrosion-resistant surface
• Easy-to-clean
• Meets demanding needs in pharmaceutical
and clinical laboratories
11
Lockable door restricts access
An audible door alarm notifies the
operator when the door is left open
Stainless steel exterior meets
demanding needs in pharmaceutical
and clinical laboratories
Unique optional sample sensor
• Measure exact sample temperature:
shown on display in addition to chamber temperature
• When used with “auto dry” program, the oven will switch off
automatically as soon as sample is dry to save energy
!
SPECIFICATIONS TABLE/ORDER NUMBERS ADVANCED PROTOCOL SECURITY OVENS SPECIFICATIONS TABLE/ORDER NUMBERS ADVANCED PROTOCOL SECURITY OVENS
Order number (coated exterior) 51028118 51028119 51028120 Order number (coated exterior) 51028127 51028128 51028129
Model OGH60-S OGH100-S OGH180-S Model OMH60-S OMH100-S OMH180-S
Order number (stainless steel exterior) 51028536 51028537 51028538 Order number (stainless steel exterior) 51028539 51028326 51028540
Model OGH60-S SS
1
OGH100-S SS
1
OGH180-S SS
1
Model OMH60-S SS
1
OMH100-S SS
1
OMH180-S SS
1
Convection technology Gravity convection Gravity convection Gravity convection Convection technology Mechanical convection Mechanical convection Mechanical convection
Temperature range °C 50 °C
2
- 330 °C 50 °C
2
- 330 °C 50 °C
2
- 330 °C Temperature range °C 50 °C
2
- 330 °C 50 °C
2
- 330 °C 50 °C
2
- 330 °C
Spatial temperature deviation at 150 °C ± 2.5 °C ± 2.5 °C ± 2.9 °C Spatial temperature deviation at 150 °C ± 1.5 °C ± 1.3 °C ± 1.5 °C
Temperature deviation over time at 150 °C ± 0.3 °C ± 0.3 °C ± 0.3 °C Temperature deviation over time at 150 °C ± 0.2 °C ± 0.2 °C ± 0.2 °C
Footprint m
2
/ sqft 0.3 / 3.2 0.36 / 3.9 0.47 / 5.1 Footprint m
2
/ sqft 0.3 / 3.2 0.36 / 3.9 0.47 / 5.1
Chamber volume L / cuft 61 / 2.15 99 / 3.6 168 / 6.05 Chamber volume L / cuft 62 / 2.2 97 / 3.5 170 / 6.1
Dimensions chamber, mm / in (W x H x D) 328 x 480 x 389 / 12.9 x 18.9 x 15.3 438 x 580 x 389 / 17.2 x 22.8 x 15.3 438 x 680 x 564 / 17.2 x 26.8 x 22.2 Dimensions chamber, mm / in (W x H x D) 354 x 508 x 343 / 13.9 x 20 x 13.5 464 x 608 x 343 / 18.3 x 23.9 x 13.5 464 x 708 x 518 / 18.3 x 27.9 x 20.4
exterior
3
, mm / in (W x H x D) 530 x 720 x 565 / 20.9 x 28.3 x 22.2 640 x 820 x 565 / 25.2 x 32.3 x 22.2 640 x 920 x 738 / 25.2 x 36.2 x 29.1 exterior
3
, mm / in (W x H x D) 530 x 720 x 565 / 20.9 x 28.3 x 22.2 640 x 820 x 565 / 25.2 x 32.3 x 22.2 640 x 920 x 738 / 25.2 x 36.2 x 29.1
Number of shelves supplied / max 2 / 13 2 / 16 2 / 19 Number of shelves supplied / max 2 / 13 2 / 16 2 / 19
Max. shelf load kg / lb 25 / 55 25 / 55 25 / 55 Max. shelf load kg / lb 25 / 55 25 / 55 25 / 55
Rated voltage / frequency V / Hz 120 / 60 208-240 / 60 208-240 / 60 Rated voltage / frequency V / Hz 120 / 60 208-240 / 60 208-240 / 60
Rated power / max. current W / A 1730 / 14.4 3360 / 14 3360 / 14 Rated power / max. current W / A 1440 / 12 3360 / 14 3360 / 14
Energy consumption at 150° C W 170 210 290 Energy consumption at 150° C W 275 389 448
Weight kg / lb 44 / 97 55 / 121 69 / 152 Weight kg / lb 44 / 97 55 / 121 69 / 152
1
Stainless Steel
2
Temperatures as low as ambient +10 °C can be selected – requires open damper and no additional heat in unit
3
Depth of handle / display not included in depth (2.6 in); adjustable feet not included in height (35 mm) – required distance to rear wall: 3.1 in
NOTE: All figures in this table are typical values for series devices, based on factory following norm DIN12880 (measured with 240V, values will differ at 208V). Please contact sales representative for certification information for IQ/OQ documents.