Operating Manual

Chapter 4 Set up Additional Measurements Accu-Wave
4-4 TN Technologies
M months
y years
Measurement Parameter and Units Menus
“Set up Additional Measurements” Menu
(Measurement Specific Parameters and Units)
Display Comments
Note: After selecting a measurement type, menu item(s) prompt for the required
parameters to finish setting up the measurement, as shown below.
If “Distance” or “% distance” are selected, the following menu items are displayed.
Distance units = ft
To change to yd, M,
cm, mm, or in press
NEXT
Available units depend on selection of “All,”
“English,” or “Metric” units under “Set up Level,…”
menus. It is possible (perhaps not often useful) to set
up distance as the primary measurement and as
another measurement number using different units.
No units are required for “% distance.”
Distance REF LINE to
zero distance point
0.000 ft
NEXT HELP
Appears if distance is not the primary measurement.
Enter the distance from the reference line marked on
the sensor to the desired zero distance point on the
tank.
REF LINE to maximum
distance to be
measured 0.000 ft
NEXT HELP
Appears if distance is not the primary measurement.
Enter the distance from the reference line marked on
the sensor to the desired maximum distance point on
the tank.
If “Level” or “% lev” are selected, the following menu items are displayed.
Level units = ft
To change to yd, M,
cm, mm, or in press
NEXT
Available units depends on selection of “All,”
“English,” or “Metric” units under “Set up Level,…”
menus. It is possible (although perhaps not often
useful) to set up level as the primary measurement
and as another measurement number using different
units. No units are required for “% lev.”
Distance REF LINE to
zero level point
0.000 ft
NEXT
Appears if level is not the primary measurement.
Enter the distance from the reference line marked on
the sensor to the desired zero level point on the tank.
Maximum level
to be measured.
0.000 ft
NEXT
Appears if level is not the primary measurement.
Enter the maximum level to be measured relative to
the zero level point.
If “Volume” or “sig str” are selected, no additional setup items are displayed. Use
the “tank volume setup” menus under “Set up level,…” to change the volume
parameters.
If “Ullage,” “% full,” or “% empty” are selected, the following additional menu item
is displayed. The units for ullage will be the same as specified for the volume.
100% Volume =
Enter the tank volume when full. The volume of the