Operating Manual
Chapter 4 Set up Additional Measurements Accu-Wave
4-2 TN Technologies
It is good practice to make a list of all the measurements you set up and
keep the list for future reference.
Select Measurement Menus
“Set up Additional Measurements” Menu
(Specify Measurement Type)
Display Comments
Set up additional
measurements
(readouts).↓↓
Other functions→→
You can assign up to 7 measurements in addition to
the primary measurement. (Primary measurement
setup is described in Chapter 3.) Press ↓ to set up
additional measurements.
NOTE: Meas. #1 is the
primary measurement
See “level setup”
to modify. NEXT↓↓
Measurement #1 is assigned to the primary
measurement by default. (Use the “Set up level,…”
menu to change the primary measurement setup as
described in Chapter 3.)
Assign & set up
measurement 2→→
NEXT↓↓
Press → to assign a readout to measurement 2. After a
readout is assigned to measurement 2, this menu item
will read “Modify setup of measurement 2” and show
the readout assigned, e.g., “cu ft” indicating volume
is assigned to measurement 2.
←←Exit measurement 2
Set up for level,
dist, %lev, %dist,
etc. NEXT↓↓
Info message only. Press ↓ to set up measurement, or
← to exit measurement 2 setup. After measurement 2
is set up, this menu item will read “Change or disable
measurement setup.”
Reading represented by
measurement 2 is
(inactive)
NEXT↓↓ CHANGE→→
Press → to select the readout for measurement 2.
Select from Distance, Level, volume, ullage, % full,
% empty, % distance, % level, sig str, mass, and rate.
You must select a measurement before other setup
menu items will be displayed. Select inactive to skip
or to deactivate a measurement.
Measurement Specific Parameters and Units
Depending on the measurement type selected you will be prompted to enter units and
certain parameters to complete the measurement setup.
Mass Measurement
If mass is selected as the measurement type, you are prompted to select the units from
among the following:
• gram grams
• kgram kilograms