Operating Manual

Accu-Wave Chapter 3 Set up Level, Level Alarms, and Volume
TN Technologies 3-5
Example: Set up Level
Set up level as the primary measurement for a tank. Before beginning the gauge setup, you
need to measure the distance from the Reference Line on the sensor housing to the desired
Zero Level point on the tank and the distance from the Zero Level point to the desired
Maximum Level to be measured. See Figure 3.3.
Figure 3.3 Entries Required to Set up Level as Primary Measurement
Follow the steps in the table below to set up level as the primary measurement using the
example measurements shown in Figure 3.3.
Example: Set up Level as Primary Measurement
Menu Display User Action
Set up level, level
alarms, & volume.
Exit Setup.
Other functions
From the measurement display, press to move to
the “Set up level,...” menu shown at the left.
Then, press to access the “Set up level,…” menu
items.
General HELP text.
Information on how
to set up this gauge
NEXT
General HELP text for setting up the gauge. Press
to read the help text, press to proceed to the next
menu item.
primary measurement:
Distance
Change to Level
NEXT
The default primary measurement is “distance.” Press
the key to change to “level.” Then press to
proceed to the next menu item.
Reference Line
Zero Level
Maximum Level
Zero Level
22.5 ft